CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by PATERSON, J. G.
Right arrow Articles by VANHOOYDONK, V.
Right arrow Search for Related Content
PubMed
Right arrow Articles by PATERSON, J. G.
Right arrow Articles by VANHOOYDONK, V.

Canadian Journal of Anesthesia, Vol 22, 373-377, Copyright © 1975 by Canadian Anesthesiologists' Society

A Hazard Associated with Improper Connection of the Bain Breathing Circuit

J. G. PATERSON 1 and V. VANHOOYDONK 1

1 Department of Anaesthesia, University of Western Ontario and Victoria Hospital, London, Ontario

A case of cardiac arrest is presented which was caused by improper connection of a modified Mapleson D circuit (Bain breathing circuit). Excessive external deadspace was created by interchanging the gas inflow line and the attachment for an airway pressure manometer. This resulted in a marked respiratory acidosis, clinically undetected until the concomitant hypoxia produced a severe cardiac depression and arrest. The sequence of events was reproduced in a dog under comparable anaesthetic conditions.

In order to avoid this error it is recommended to permanently fuse the connecting piece placed in the circuit with the tubing leading to the airway pressure manometer of the respirator.