Canadian Journal of Anesthesia, Vol 23, 135-142, Copyright © 1976 by Canadian Anesthesiologists' Society

Magill Circuit and Controlled Ventilation

M. O. OBIAYA 1 and P. DAKARAJU 1

1 Department of Anaesthesiology, University of Benin, Benin City, Nigeria

Nine patients under anaesthesia were controlled manually with the Magill attachment. Nine other patients using the same circuit under anaesthesia breathed spontaneously. Blood gases were studied throughout the period of operation to determine the adequacy of ventilation. The results show that the Magill attachment was adequate for controlled ventilation using normal flows (9 l/min) in young healthy patients.