CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by PING, F. C.
Right arrow Articles by JENKINS, L. C.
Right arrow Search for Related Content
PubMed
Right arrow Articles by PING, F. C.
Right arrow Articles by JENKINS, L. C.

Canadian Journal of Anesthesia, Vol 26, 415-419, Copyright © 1979 by Canadian Anesthesiologists' Society

Bacterial Filters-Are they Necessary on Anaesthetic Machines?

F. C. PING M.B., Ch.B., B.A.O.1, J. L. OULTON M.D., C.M., F.R.C.P.(C).1, J. A. SMITH M.D., F.R.C.P.(C)1, A. G. SKIDMORE M.D., F.R.C.P.(C)1, and L. C. JENKINS M.D., C.M., F.R.C.P.(C)1

1 Department of Anaesthesia and Division of Microbiology, University of British Columbia and Vancouver General Hospital, Vancouver. B.C.

At the Vancouver General Hospital the effectiveness of the system for decontamination of anaesthetic equipment was evaluated to determine the need for bacterial filters on anaesthetic machines.

Two groups of patients were studied. Group I consisted of 33 patients, none of whom had clinical symptoms of respiratory tract disease. Group II consisted of 17 patients who had lower respiratory tract secretions. In the latter group 16 had chronic bronchitis and one had cystic fibrosis.

Of 550 bacterial cultures taken from the anaesthetic equipment immediately before and after anaesthesia in our 50 patients, only five yielded a growth of non-pathogenic bacteria.

The results of this study indicate that bacterial colonization of anaesthetic equipment is of a low order and is adequately controlled by pasteurization even after use in patients with chronic lower respiratory tract disease. The use of bacterial filters does not appear justified if a strict regimen of cleaning and pasteurization is followed.







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS