| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Canadian Journal of Anesthesia, Vol 38, 985-988, Copyright © 1991 by Canadian Anesthesiologists' Society
ARTICLES |
I Dimich, PP Singh, A Adell, M Hendler, N Sonnenklar and M Jhaveri
Department of Anesthesiology, Mount Sinai Medical School, New York, New York.
The pulse oximeter was evaluated for use in neonates in the delivery room. One hundred neonates, delivered vaginally or by Caesarean section with general or epidural anaesthesia, were studied. After delivery, pulse oximetry probes were placed simultaneously on the ulnar side of the right hand and on the right Achilles tendon to determine whether there was a difference in arterial oxygenation (SpO2). Measurements of SpO2 were taken at 1, 5, 10 min, and 24 hr after delivery. At one and five minutes, SpO2 recorded from the right hand was higher than that recorded from the lower extremities (71.9% +/- 6.5% vs 63.4% +/- 4.3% and 83.3% +/- 4.2% vs 76% +/- 4.1%, mean +/- SD, respectively). At ten minutes these differences diminished, and had almost completely disappeared after 24 hr. These results can be explained by the presence of R-L shunting at the ductus arteriosus level, producing reduced SaO2 in the lower extremities. Oxygen saturation did not differ between neonates delivered vaginally or by Caesarean section, regardless of the presence or type of anaesthesia. We concluded that neonates remain relatively desaturated in the immediate postpartum period and that the SpO2 obtained from the right hand is a better index of neonatal oxygenation than that obtained from the heel.
This article has been cited by other articles:
![]() |
J A Dawson, P G Davis, C P F O'Donnell, C O F Kamlin, and C J Morley Pulse oximetry for monitoring infants in the delivery room: a review Arch. Dis. Child. Fetal Neonatal Ed., January 1, 2007; 92(1): F4 - F7. [Full Text] [PDF] |
||||
![]() |
L M O'Brien, V A Stebbens, C F Poets, E G Heycock, and D P Southall Oxygen saturation during the first 24 hours of life Arch. Dis. Child. Fetal Neonatal Ed., July 1, 2000; 83(1): 35F - 38. [Abstract] [Full Text] |
||||
![]() |
V. K. Reddy, I. R. Holzman, and J. F. Wedgwood Pulse Oximetry Saturations in the First 6 Hours of Life in Normal Term Infants Clinical Pediatrics, March 1, 1999; 38(2): 87 - 92. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |