CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chartrand, D.
Right arrow Articles by Lapointe, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chartrand, D.
Right arrow Articles by Lapointe, A.

Canadian Journal of Anesthesia, Vol 40, 1076-1083, Copyright © 1993 by Canadian Anesthesiologists' Society


ARTICLES

Measurement of respiratory mechanics using the Puritan-Bennett 7200a ventilator

D Chartrand, B Dionne, C Jodoin, M Lorange and A Lapointe
Department of Anaesthesia, Hotel-Dieu de Montreal, University of Montreal, Quebec.

This study was designed in order to validate the respiratory mechanical variables measured by the Puritan-Bennett 7200a ventilator equipped with the 30/40 module. Two ventilators were connected to a lung model and submitted to several breathing patterns by modifying the respiratory rate, the tidal volume, the inspiratory flow-rate and the model resistance. The inspiratory flow-rate (V), tidal volume (VT), peak inspiratory pressure (Pmax), plateau pressure (Pplat) and PEEP measured by the ventilators were compared with the same variables measured at the connection between the breathing circuit and the lung model. The compliance (C30/40) and the resistance (R30/40) calculated by the 30/40 module were compared with those calculated by using the variables measured by the reference equipment. Both ventilators made a constant underestimation of V by 2.8 and 3.7 L.min-1, respectively. The VT was measured with a mean error of less than 10 ml but did not reflect the preselected values in the presence of an intrinsic PEEP. The Pplat was overestimated by 7 and 10%, respectively. The same calibration error was observed with Pmax which was also affected by a pressure gradient due to the resistance of the breathing circuit. Even in the absence of intrinsic PEEP, C30/40 presented an error due to the combination of the measurement errors on VT, Pplat and PEEP. Finally, R30/40 presented a high percentage of error due to the combination of the measurement errors on V, Pmax, and Pplat, and to a sporadic aberrant selection of V. Due to these numerous sources of error, the two ventilators studied did not give reliable estimates of resistance and compliance.(ABSTRACT TRUNCATED AT 250 WORDS)