CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Searle, N. R.
Right arrow Articles by Dupont, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Searle, N. R.
Right arrow Articles by Dupont, C.

Canadian Journal of Anesthesia, Vol 40, 388-393, Copyright © 1993 by Canadian Anesthesiologists' Society


ARTICLES

Assessment of the arterial tonometer (N-CAT) for the continuous blood pressure measurement in rapid atrial fibrillation

NR Searle, J Perrault, H Ste-Marie and C Dupont
Department of Anesthesiology, Montreal Heart Institute, Quebec, Canada.

The N-CAT is a newly developed arterial tonometer (TBP) able to determine systolic, diastolic and mean arterial blood pressures continuously and noninvasively. The aim of this study was to evaluate the accuracy and reliability of TBP relative to directly measured invasive blood pressure (IBP) in ten haemodynamically stable postoperative cardiac patients who were in rapid atrial fibrillation (HR > or = 100 bpm). There were differences between TBP and IBP for systolic (-1.7 mmHg) and diastolic (+0.9 mmHg) values but not for the mean arterial blood pressures. The N-CAT was able to follow blood pressure changes closely and demonstrated an average systolic, diastolic and mean bias (+/-SD) of -1.71 +/- 4.6, 0.99 +/- 4.6 and 0.33 +/- 4.2 mmHg, respectively. Although these biases are within the required standards for equivalency for noninvasive blood pressure to invasively determined blood pressure, approximately 20% of the readings were > +/- 10 mmHg while only 5% were > +/- 20 mmHg. Moreover, there were occasional discrepancies of sufficient magnitude and duration which may limit the clinical usefulness of the N-CAT in patients in whom continuous and accurate blood pressure measurement is required.


This article has been cited by other articles:


Home page
Anesth. Analg.Home page
G. M. Janelle and N. Gravenstein
An Accuracy Evaluation of the T-Line(R) Tensymeter (Continuous Noninvasive Blood Pressure Management Device) versus Conventional Invasive Radial Artery Monitoring in Surgical Patients
Anesth. Analg., February 1, 2006; 102(2): 484 - 490.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1993 by the Canadian Anesthesiologists' Society.