JD Tobias, DJ Meyer Jr and MA Helikson
Department of Child Health, University of Missouri, Columbia 65212, USA.
PURPOSE: To determine the accuracy of the Paratrend 7 monitor inserted into a peripheral vein in providing continuous measurements of PCO2 and pH in children with respiratory failure. METHODS: In a prospective, case series at a University based PICU in a tertiary referral hospital and following informed consent, the Paratrend 7 intra-arterial blood gas monitoring system was inserted through a peripheral intravenous cannula. When clinically indicated, an arterial blood gas (ABG) was obtained. The values of pH and PCO2 from the Paratrend 7 and the ABG were compared using linear regression analysis and Bland-Altman analysis. RESULTS: Seventeen ABGs were obtained in the 4 patients. PaCO2 ranged from 29 to 51 mmHg and the pH values from 7.16 to 7.50. Linear regression analysis for the: Paratrend 7 pH vs the ABG pH resulted in an r = 0.9678 and an r2 = 0.9367 (P < 0.0001). Linear regression analysis for the Paratrend 7 CO2 vs ABG CO2 resulted in an r = 0.8273 and an r2 = 0.6845 (P = 0.01). Using a Bland and Altman analysis, the bias +/- precision of the PCO2 values was 3 +/- 1.4 and 0.036 +/- 0.02 for the pH values. CONCLUSION: The pH and PCO2 values obtained from the Paratrend 7 monitor provide a clinically acceptable correlation with arterial values.
This article has been cited by other articles:
![]() |
M. Ganter and A. Zollinger Continuous intravascular blood gas monitoring: development, current techniques, and clinical use of a commercial device Br. J. Anaesth., September 1, 2003; 91(3): 397 - 407. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Mantha, M. F. Roizen, L. A. Fleisher, R. Thisted, and J. Foss Comparing Methods of Clinical Measurement: Reporting Standards for Bland and Altman Analysis Anesth. Analg., March 1, 2000; 90(3): 593 - 602. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. L. Williams, K. L. Hildebrand, S. A. McCormick, and M. J. Bedel The Effect of Intravenous Lactated Ringer's Solution Versus 0.9% Sodium Chloride Solution on Serum Osmolality in Human Volunteers Anesth. Analg., May 1, 1999; 88(5): 999 - 1003. [Abstract] [Full Text] [PDF] |
||||