CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Résumé de cet Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Araki, K.
Right arrow Articles by Hatano, Y.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Araki, K.
Right arrow Articles by Hatano, Y.
Related Collections
Right arrow Cardiothoracic Anesthesia, Respiration and Airway
Canadian Journal of Anesthesia 47:775-779 (2000)
© Canadian Anesthesiologists' Society, 2000

Reports of Investigation

Bronchial cuff pressure change caused by left-sided double-lumen endobronchial tube displacement

Kazukuni Araki, MD*, Ryoichi Nomura, MD*, Reiko Urushibara, MD*, Yukiko Yoshikawa, MD* and Yoshio Hatano, MD{dagger}

* From the Departments of Anesthesiology, Ohtsu Red Cross Hospital, 1-1-35 Nagara, Ohtsu, Shiga 520-8511, Japan; and
{dagger} Wakayama Medical College, Wakayama, Japan.

Address correspondence to: Address correspondence to: K. Araki MD. Phone: 81-77-522-4131; Fax: 81-77-525-8018.

Purpose: The bronchial cuff pressures (BCPs) of left-sided double-lumen endobronchial tubes (DLTs) manufactured by Rüsch and Mallinckrodt were measured in 80 patients when the tubes were withdrawn to compare the effect of tube design on BCP change.

Methods: During general anesthesia with muscle relaxation, the cephalad surface of the endobronchial cuff was positioned either 2.5 cm distal to the carina (Rüsch Group R-I; n = 20 and Mallinckrodt Group B-I; n = 20) or just below the carina (Rüsch Group R- II; n = 20 and Mallinckrodt Group B- II; n = 20) and the cuff was inflated to 35 cm H2O. The tube was then withdrawn in 0.5-cm steps until the cuff was 2.0 cm proximal to the carina, the position just before the capnogram or pressure-volume loop of tracheal lumen changed. The BCP at each step was measured. The rate of decrease in BCP was defined as the decrease of BCP divided by the length of displacement of DLT.

Results: The rates of decrease from the +2.5 cm position to the end point in Group B-I (7.7 ± 0.8 cm H2O•cm-1) and those from the most proximal acceptable position to the end point in Group B-II (19.5 ± 4.8 cm H2O•cm-1) were greater than those in Group R-I (6.9 ± 0.9 cm H2O•cm-1) (P < 0.01) and in Group R-II (12.4 ± 3.1 cm H2O•cm-1) (P < 0.01), respectively.

Conclusion: The BCP decreased in both of the Mallinckrodt and Rüsch DLTs, and the rates of decrease of the former were greater than those of the latter.




This article has been cited by other articles:


Home page
SEMIN CARDIOTHORAC VASC ANESTHHome page
E. Mirzabeigi, C. Johnson, and A. Ternian
One-Lung Anesthesia Update
Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2005; 9(3): 213 - 226.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2000 by the Canadian Anesthesiologists' Society.