| Abstract |
|---|
|
|
|---|
Methods: An intubation mannequin head was modified so that, using a Macintosh blade size 3, only the epiglottis was visible at direct laryngoscopy, representing a grade III laryngoscopic view. Forty anesthesiologists attempted tracheal intubation using each technique. Tracheal intubation with the Bullard laryngoscope was performed using the attached non-malleable intubating stylet preloaded with an endotracheal tube. The video-optical intubation stylet inserted into an endotracheal tube was used with direct laryngoscopy. During conventional laryngoscopy, the video-view from the stylet tip allowed the tracheal tube to be guided behind the epiglottis into the trachea.
Ten attempts with each technique were performed by each anesthesiologist in randomized order. Intubation time, and failed intubation (> 60 sec / esophageal intubation) were recorded. The operators assessed the degree of difficulty of each method using a Likert-scale.
Results: Mean intubation time (19.2 ± 4.5 sec for the BL and 18.8 ± 4.6 sec for the VOIS) was almost the identical. The video-optical intubation stylet was associated with fewer failed intubations (8 vs 41; P < 0.005) and had a lower degree of difficulty (1.7 ± 0.65 for the VOIS and 2.6 ± 0.74 for the BL; P < 0.0001). No correlation was found between the anesthesiologist's experience and mean intubation time, estimated degree of difficulty or number of unsuccessful intubation.
Conclusion: The video-optical intubation stylet was a more effective and simpler intubation device to facilitate difficult tracheal intubation than the Bullard laryngoscope.
| Introduction |
|---|
|
|
|---|
The Bullard laryngoscope (BL) (Circon ACMI, Stamford, CT), is a rigid, fibreoptic intubating laryngoscope for indirect oral laryngoscopy, transmitting the view from the tip of the L-shaped blade to a proximal view finder.4 It has been reported to be a valuable aid for the management of the predicted as well the unanticipated difficult tracheal intubation.5,6
The video-optical intubation stylet (VOIS) (VOLPI AG, Schlieren, Switzerland) is a new intubation device, especially designed for rapid management of the unexpected difficult intubation.7 Inserted into an endotracheal tube (ETT), the VOIS transmits the view from the stylet tip onto a bedside monitor. During difficult direct laryngoscopy, this allows the anesthesiologist to guide the endotracheal tube into the trachea using the video-view from the stylet tip.7,8
The purpose of this study was to compare the ease and success of tracheal intubation facilitated by the video-optical intubation stylet or the Bullard laryngoscope in a model of difficult intubation.
| Methods |
|---|
|
|
|---|
Forty anesthesiologists of different years of experience and without experience with the intubation device, were asked to attempt tracheal intubation using each method:
Bullard laryngoscope using the non-malleable intubating stylet
Before starting intubation an ETT I.D. 7.0 mm with cuff (RUESCHELITTM, Willy Ruesch AG, Kernen, Germany) was loaded on the non-malleable intubation stylet attached to the BL. Tracheal intubation with the Bullard laryngoscope was performed as described and evaluated by Cooper et al.10
Video-optical intubation stylet inserted into an endotracheal tube
A malleable video-optical intubation stylet (Volpi AG, Schlieren, Switzerland) with O.D. 5 mm was used for tracheal intubation.7 The malleable intubation stylet has integrated 10,000 image fibres and additional light transmitting fibres. They leave the stylet at it's proximal end as a 1.7 metre long ultra-thin cable (O.D. 2.8 mm) which is attached to a video-monitor system. The VOIS was inserted into an ETT I.D. 7.0 mm with cuff (RUESCHELITTM, Willy Ruesch AG, Kernen, Germany) slightly protruding the ETT tip.
Conventional laryngoscopy was performed by the operator as usual. While maintaining the best laryngoscopic view (grade III only possible) the distal end of the ETT, bent into an appropriate hockey stick curve, was placed by the operator behind the epiglottis under direct vision. Then, using the monitor view from the stylet tip, the ETT was directed through the glottic opening into the trachea. Final tracheal ETT position was confirmed respectively adjusted using the monitor view before removing the stylet from the ETT.
After one demonstration of each intubation technique by the supervisor, the operator was allowed to do one practice intubation attempt with each technique. Then, each anesthesiologist performed each technique 10 times in succession (order randomized). The time was recorded for each attempt from initial insertion of the blade into the oral cavity until complete removal of the intubation device from the ETT. Failed intubation was defined as requiring > 60 sec or which resulted in esophageal intubation. At the end of each sequence, the operators were asked to attach a degree of difficulty to the method using a Likert-scale from 1 (extremely simple technique) to 5 (impossible technique).
The results were statistically evaluated using Mann-Whitney-U-test for comparison of intubation times and Chi2- test for comparison of proportions of failure rates and degrees of difficulty. A P value of < 0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
In the present study, we compared the ease and success of tracheal intubation using the Bullard laryngoscope or the video-optical intubation stylet in a difficult intubation setting. The results showed the superiority of the VOIS to the BL regarding success rates and degree of estimated difficulty of intubation but not of mean intubation time.
Failed intubation was more common with the Bullard laryngoscope. The higher esophageal intubation rate in the Bullard group can be explained by the ETT obstructing the endoscopic view of the cords during intubation. During tracheal intubation with the VOIS, the view of the cords is never obstructed and tracheal intubation is immediately confirmed by recognizing the tracheal rings. The only recorded esophageal intubation in the VOIS group was immediately recognized by the operator on the video-display but intubation time was already > 60 sec. The video-optical intubation stylet allows recognition and correction of esophageal intubation during tube insertion, before gastric air insufflation occurs. In addition, the view from the ETT tip helps in the avoidance of accidential endobronchial intubation.
Most of the intubations were performed within an acceptable time with either technique. Since the Bullard laryngoscope has been equipped with a non-malleable stylet, tracheal intubation with the BL has become simple and rapid.10 Although the VOIS technique includes two maneuvres such as direct laryngoscopy and stearing a tracheal tube according to the video-view, mean intubation times were similar in both groups. Comparison of failed intubation at 30 sec and 45 sec was not different between the two techniques (Table I
), indicating that both techniques are suitable endoscopic aids in patients with limited cardio-respiratory reserves.
Using the video-optical intubation stylet in a patient with a grade III laryngoscopic view is similar to using a gum elastic bougie or stylet. However, instead of blind probing for the glottic opening, the endotracheal tube is directed into the trachea using the video-view from the stylet tip. The familar nature of the VOIS technique was reflected by the decreased estimate of difficulty particularly in the group of experienced anesthesiologists, the higher success rate in the first five intubation attempts compared with the BL and the lack of a learning curve in the VOIS (failed intubation).
The video-optical intubation stylet can be used with a standard intubation technique. There is no need to change the intubation procedure if an unanticipated difficult laryngoscopy occurs. When the VOIS has been already inserted into the ETT before starting a rapid sequence induction, it does not interfere with conventional tracheal intubation, but can be used instantly when urgent endoscopic intubation assistance is required.
| Conclusion |
|---|
|
|
|---|
Accepted for publication November 26, 1999.
| References |
|---|
|
|
|---|
2 Cheney FW. The American Society of Anesthesiologists closed claims project. What have we learned, how has it affected practice, and how will it affect practice in the future? Anesthesiology 1999; 91: 5526.[Medline]
3 Tse JC, Rimm EB, Hussain A. Predicting difficult endotracheal intubation in surgical patients scheduled for general anesthesia: a prospective blind study. Anesth Analg 1995; 81: 2548.[Abstract]
4 Bjoraker DG.The Bullard intubating laryngoscopes. Anesthesiology Review 1990; 17: 6470.[Medline]
5 Gorback MS. Management of the challenging airway with the Bullard laryngoscope. J Clin Anesth 1991; 3: 4737.[Medline]
6 Kiefer G, Hentrich B. The Bullard laryngoscope: help when unforseen difficulties with intubation arise (German). Anästhesist 1996; 45: 704.[Medline]
7
Weiss M. Video-Intuboscopy: a new aid to routine and difficult tracheal intubation. Br J Anaesth 1998; 80: 5257.
8 Weiss M. Management of difficult tracheal intubation with a video-optically modified Schroeder intubation stylet (Letter). Anesth Analg 1997; 85: 11812.[Medline]
9 Cormack RS, Lehane J. Difficult tracheal intubation in obstetrics. Anaesthesia 1984; 39: 110511.[Medline]
10
Cooper SD, Benumof JL, Ozaki GT. Evaluation of the Bullard Laryngoscope using the new intubating stylet: comparison with conventional laryngoscopy. Anesth Analg 1994; 79: 96570.
11 Goldberg JS, Bernard AC, Marks RJ, Sladen RN. Simulation technique for difficult intubation: teaching tool or new hazard? J Clin Anesth 1990; 2: 216.[Medline]
12
Hodges UM, O'Flaherty D, Adams AP. Tracheal intubation in a mannikin: comparison of the Belscope with the Macintosh laryngoscope. Br J Anaesth 1993; 71: 9057.
This article has been cited by other articles:
![]() |
O. Hung and J. A. Law Advances in airway management Can J Anesth, June 1, 2006; 53(6): 628 - 631. [Full Text] [PDF] |
||||
![]() |
E. B. Liem, D. G. Bjoraker, and D. Gravenstein New options for airway management: intubating fibreoptic stylets{dagger} Br. J. Anaesth., September 1, 2003; 91(3): 408 - 418. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Dullenkopf, G. Lamesic, A. Gerber, and M. Weiss Video-enhanced visualization of the larynx and intubation with the Bullard laryngoscope - equipment report: [La visualisation, facilitee par video, du larynx et de l'intubation avec le laryngoscope Bullard - une analyse de materiel] Can J Anesth, May 1, 2003; 50(5): 507 - 510. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. B.P. Stasiuk Improving styletted oral tracheal intubation: rational use of the OTSU : [L'amelioration de l'intubation oro-tracheale : l'utilisation rationnelle du SIOT] Can J Anesth, October 1, 2001; 48(9): 911 - 918. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Weiss, K. Hartmann, J. Fischer, and A. C. Gerber Video-intuboscopic assistance is a useful aid to tracheal intubation in pediatric patients : [L'assistance videoscopique facilite l'intubation endotracheale des patients pediatriques] Can J Anesth, July 1, 2001; 48(7): 691 - 696. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Weiss, U Schwarz, C. Dillier, and A. Gerber Video-intuboscopic monitoring of tracheal intubation in pediatric patients Can J Anesth, December 1, 2000; 47(12): 1202 - 1206. [Abstract] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |