| Abstract |
|---|
|
|
|---|
Source: CSA-International, International Electrotechnical Committee (IEC), Organization for International Standardization (ISO).
Principal findings: The safety record of the practice of anesthesia has improved over the past 20 yr. This has been brought about by the increase in the knowledge base of physiology and pharmacology, improvements in anesthetic drugs and other medications and through the use of better equipment. However, the introduction of new equipment and the building of new health care facilities has also introduced risks to patient safety. These risks have been created by oversights in design, mistakes in equipment manufacture or facility construction and the need to produce a usable product for the price the health care system can afford. After the anesthetic disasters of the 1960s and 1970s, the need for minimum standards for safety in the design and construction of healthcare facilities and medical equipment were recognized. Canadian anesthesiologists were at the forefront in the development of these standards, and still today members of the Canadian Anesthesiologist's Society are key contributors to the development of both National and International standards.
Conclusion: Equipment and facility standards have helped improve patient safety over the past 20 yr. However, anesthesiologists need to encourage their governmental regulatory bodies to encourage manufacturer compliance with specific standards.
ONE morning in August 1973, a six-year-old girl fell from a swing at her grandmother's home and badly fractured her leg. She was taken to the hospital 50 km away, where she died during surgery. Five months later, a coroner's jury ruled that the girl had died "due to administration of nitrous oxide following an anesthetic." A fatal error in the hospital's medical gas pipeline system resulted in the administration of nitrous oxide 100% during emergence instead of oxygen 100%. The jury also found that the same problem had either caused or contributed to eight other deaths at the hospital in 1973.1
A 53-yr-old woman entered a hospital to have a bunion removed from her toe. A coroner's jury later found that an improperly installed valve in the anesthetic machine caused anoxic encephalopathy that led to her death.1
There are many different types of standards in the world today. Everyone has noticed the CSA mark on the back of household appliances, stereo systems, children's bicycle helmets and anesthetic and critical care equipment. Anesthesiologists also have standards of practice for patient care, but what are these standards about, who writes them, and what do they mean in our daily practice of medicine? In this article, I will describe the purpose of equipment standards, how they are written, how they should be applied, their legal status and address the current issues surrounding anesthesia, respiratory care and critical care equipment standards, their advantages and limitations.
| Definitions |
|---|
|
|
|---|
Within Anesthesia and Critical Care there are three types of standards:
| Standard development |
|---|
|
|
|---|
Initial drafts are written by working groups or adapted from similar standards from other countries or from standards of similar equipment. Members from national committees are asked to comment and vote on these drafts. The normal progression of the drafts include working drafts to which comments and corrections can be made, draft standards for comment, and then final drafts which are voted on to accept or reject only. (Figure 1
)
|
Most of the anesthesia and critical care standards are written by the Organization for International Standardization ISO Technical Committee 121 Anesthesia and Respiratory Equipment (ISO TC121) and its subcommittees, International Electrotechnical Commission IEC Committee 62 and in Canada, the CSA-International Committee Z251 (Table I
, Figure 2
). ISO TC 121 first met in London UK in 19672 and adopted the British standard of 15 mm and 22 mm for breathing system connections. The Canadian Anesthesiologists' Society has representatives on the CSA International, ISO and IEC Committees. These anesthesiologists bring an important clinical perspective to these committees that are largely made up of engineers from the various equipment manufacturers. The opinions of anesthesiologists and intensivists are very highly regarded by the manufacturers who have little practical experience with equipment within the operating room and critical care settings.
|
|
| Who are the standards bodies and organizations? |
|---|
|
|
|---|
There are numerous standards writing committees and organizations throughout the world (Table II
). As anesthesiologists in Canada, we use the standards of the CSA International, the Organization for International Standardization (ISO), the International Electrotechnical Commission (IEC), and the Compressed Gas Association (CGA). In the United States the anesthetic and critical care Committee F29 of the American Society for Testing and Materials (ASTM) writes the standards. In Europe the standards body is known as CEN and their marking is CE. In Japan, their standards body is the Japanese Industrial Standards Committee (JISC).
|
The International Electrotechnical Commission (IEC) has 42 member countries that account for 80% of the world's population and 95% of its electrical energy production. Members are drawn from the principal standardization bodies at the national level. The IEC maintains advisory committees on electromagnetic compatibility, electrical, medical, and telecommunications electronics safety.
CSA International
The best known standards organization in Canada is CSA International. CSA International (formerly the Canadian Standards Association) is a non-governmental, private sector, not for profit corporation with over 8,000 members and a network of offices worldwide. CSA International provides two functions: (1) to sponsor and facilitate the development of standards for products and services, and (2) as a test house to test individual pieces of equipment or certify business management practices to ascertain whether they comply with their respective standards. QMI, a wholly owned subsidiary specializes in the management systems registration for the ISO 9000 series of standards. CSA-International is a global leader in development and certification of equipment standards. In addition, they are in partnership with more than 30 international standards organizations. They are a recognized certification body for the IEC, maintain their own electromagnetic compatibility (EMC) laboratory, and have an explosive atmospheres lab. The EMC laboratory, for example, could test for possible problems with medical equipment in close proximity to cellular telephones.
The first CSA standard in the health care area was published in 1963 to deal with the problem of explosions in hospital operating rooms involving flammable anesthetic agents. The CSA Z32-1963 Code for Use of Flammable Anesthetics effectively reduced the principal hazard (involving medical equipment) of that era. Other standards followed: a standard for endotracheal tubes in 1963, for endotracheal tube connectors in 1967, the Code for Prevention of Explosions and Electric Shock in Hospital Operating Rooms in 1970, and a standard for non-flammable medical gas piping systems in 1973.
A major impetus to the expansion of health care standards activities was the report, Clinical Laboratory and Electromedical Equipment Study issued in 1973 by CSA and the Federal Department of Health. The study confirmed that doctors and hospitals were experiencing difficulties in coping with the profusion of medical devices. One major direct result of the 1973 study on laboratory and electromedical equipment was the establishment that year of CSA's Health Care Technology program.1 The technical drawings for the standard anesthesia machine oxygen knob (Figure 3
) and endotracheal tubes (Figure 4
) are examples of equipment that are very familiar to anesthesiologists.
|
|
Reading standards
Standards, like many legal documents, have a language of their own. The term "shall" denotes a requirement and means that compliance with a requirement is mandatory. All "shall" clauses are followed by a test to check for compliance. Statements using the term "should" designates that compliance with a requirement or a test is recommended but not mandatory for compliance with the standard. The term "may" is used to describe the permissible way to achieve compliance with a requirement or a test. Often written into the standards are requirements for the hospitals and for the people actually using the equipment. The term "user" is defined as the authority responsible for the use and maintenance of the equipment whereas the "operator" is the person actually handling and operating the equipment. In the case of home care equipment the user and the operator may be the same person.
Certification
Before sale, all manufacturers submit their equipment to "test houses" for certification of their equipment's performance to a particular standard. These "test houses" include but are not limited to CSA International, Underwriters Laboratories, ETL, and Entela.
Advantages of standardization
Standards attempt to improve the health and safety of the anesthesiologist, other healthcare providers and the patient. They attempt to enhance equipment quality and reliability at a price that the health care system can afford. Standards facilitate training of anesthesiologists, as a particular type of equipment is similar between models and manufacturers. Standards provide greater compatibility and interoperability of equipment across different models and different manufacturers and hopefully, standards lead to simplification of equipment for improved usability and ease of maintenance.
In the growing global economy, international standards facilitate trade between countries. As the world shrinks with the internationalization of trade, there is a need for global harmonization of existing standards. As Canada is a relatively small country, it is in our best interests to harmonize our national standards with international standards, as either manufacturers will not specifically design and build equipment for sale only in the small Canadian market or we will be unable to afford the price of Canadianized equipment.
Shortfalls of standards
The greatest disadvantage of standards is that they specify only minimum requirements for safety. The major limitation of standards, however, is that in Canada and many other countries, manufacturer compliance with existing standards is voluntary when designing equipment. It is up to the manufacturer to decide whether they wish to apply each of the applicable standards for each piece of equipment. Manufacturers are not required by law to manufacture equipment compliant to standards written for particular pieces of equipment unless mandated by a provincial or national electrical code. The Canadian electrical code requires compliance with CSA 22.2 Number 125, or CSA C22.2 No. 601.1. Number 125 is slowly being phased out and No. 601.1 will take its place. It is the Canadian adaptation of IEC 60601-1 usually referred to as the "General Standard" and provides electrical safety requirements, tip-over tests, spill tests and other safety related requirements. Ethical equipment manufacturers usually meet and exceed the requirements of these standards.
Recent issues involving standards
There is some confusion as to what the CSA mark on the back of a piece of equipment means. It is not a requirement that the manufacturer lists the standard numbers to which the equipment was tested on the sticker. It is the responsibility of the purchaser of equipment to ask the manufacturer to which standards the equipment is compliant. All manufacturers within their accompanying documents must state to which standards their equipment is compliant, but they often do not. Usually, the CSA sticker only represents compliance with the Canadian electrical code.
For example, in the case of an older conventional anesthetic machine, the machine must meet the requirements of CSA 22.2 No. 601.1 or No.125 but does not legally need to meet the requirements of CSA Z168.3 Continuous-Flow Inhalation Anaesthetic Apparatus (Anaesthetic Machines) for Medical Use.
The purchaser of equipment, however, must also be aware that not all pieces of equipment have particular performance standards. There is no Canadian standard for the new Anesthesia workstations. "Anesthetic workstation" is the generic term for equipment that delivers anesthetic agents, has an integrated ventilator and physiological multiparameter monitors. The major advantage of these workstations is that they are well integrated units with fewer hoses and wires dangling from them, and all the alarms are located in one location. Workstations facilitate automated record-keeping as all the data are from one source. Some of these anesthetic workstations have been designed to currently available IEC, European CEN and American ASTM standards.
In addition to the anesthetic workstation standard, the workstation should also be compliant with a number of other standards. These include, but are not limited to those in Table III
.
|
Equipment can only comply with the standards that are available at the time of their design. As it takes several years to design equipment and for the regulatory agencies to approve it for use, when the equipment is ultimately available for use, the standard to which it is compliant may no longer be the current edition of the standard. Similarly, as standards take several years to develop, the rapidly evolving technology may often overtake them. Thus standards' writers must be cognizant of the need to write standards that will not stifle innovation.
Over the years, each nation has developed its own standards. Until recently, anesthesia and critical care equipment produced by a single manufacturer had to adhere to the standards of all the countries in which the equipment was distributed. The proliferation of incompatible national standards greatly increases the cost of equipment. For example, the colour code for oxygen is white in Canada, green in the United States and blue in Germany! Fortunately for Canada, we already are using the ISO 32 white colour code for oxygen. Since the creation of the European Economic Community and the North American Free Trade Agreement, global standards are now being harmonized so that manufacturers in the future will only have to manufacture equipment to one international standard.
For example, there are three standards worldwide for anesthetic workstations and critical care ventilators. A joint working group of ISO, IEC and CEN is working to harmonize standards to replace these six standards with two standards. As there are limited resources in Canada for producing a new anesthetic workstation standard, it is probably in our best interest to wait for a new international standard to be developed. The Canadian opinions, however, are not lost as there are Canadian representatives on these international joint-working groups.
To try to simplify and minimize the confusion of multiple standards, the CSA-International Committee Z251 meets on a regular basis to review international standards and advises the SCC on which standards to adopt as Canadian National Standards. A listing of all the CSA standards for Anesthesia and Critical Care equipment is provided in the Canadian Anesthesiologists' Society Annual Membership List and Standards of Care edition as well as on the member's section of the CAS Website.
Standards have made anesthesia and critical care equipment much safer over the years. All hospitals should own copies of the CSA-International, ISO or IEC standards for the equipment that the hospital utilizes. These standards may be purchased from CSA-International, the Standards Council of Canada or directly from the ISO or IEC. Some standards are available as Adobe Acrobat portable document format (pdf) files and may be purchased and downloaded from the web sites listed in Table II
.
It is the responsibility of the purchaser of equipment to be aware of the value and limitations of standards on the equipment that they purchase. They should be clear about to which standards the prospective piece of equipment is compliant. When purchasing equipment, the purchase order should state that the equipment shall meet the requirements of the Canadian or provincial electrical codes and other particular equipment standards if available and if applicable. Only by requiring compliance with standards, will manufacturers be forced to comply with them. The Canadian Anesthesiologists' Society and CSA-International representatives are willing to discuss any concerns that Anesthesiologists, Intensivists, Respiratory Therapists or others have about the design and safety of equipment in Canada.
Building awareness and understanding of standardization and the important contribution it makes to Canadians' social and economic well-being is critical to the long-term success ... of Canada's standardization efforts. An informed public, with a solid understanding of the benefits and appropriate use of standardization processes, will help to ensure that decision-makers in both government and industry give full consideration to standardization as an important tool for shaping policies that meet the needs of the Canadian public and industry".5
| Acknowledgments |
|---|
Accepted for publication October 1, 2000.
| References |
|---|
|
|
|---|
2 Greenbaum R. The breath of Life. ISO/TC 121, Anaesthetic and respiratory equipment, breathes optimism and reaps results. ISO Bulletin, January 1997; 28: 114.
3 Council Directive 93/42/EEC of 14 June 1993, concerning medical devices. Official Journal of the European Communities 1993; L169: 143.
4 International Electrotechnical Commission. Medical electrical equipment. Part 1: General requirements for safety. 2. Collateral standard: electromagnetic compatibility requirements and tests. IEC 60601-1-2.
5 Standards Council of Canada, Canadian Standards Strategy. December, 1999.
This article has been cited by other articles:
![]() |
S. Dain Towards international standardization of the anesthesia machine/Vers une standardisation internationale de la plate-forme d'anesthesie Can J Anesth, August 1, 2007; 54(8): 613 - 619. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||