| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |


* From the Department of Anaesthesia, University Childrens Hospital Zurich, Switzerland
the Department of Anaesthesia, University Hospital Zurich, Switzerland
and the Research Department, BioMIP, Schloss Birlinghoven, St. Augustin, Germany
Address correspondence to: Dr. Markus Weiss, Department of Anaesthesia, University Childrens Hospital, Steinwiesstrasse 75, CH-8032 Zurich, Switzerland. Phone: +41 1 266 77 53; Fax:+41 1 266 71 68; E-mail: markus.weiss{at}kispi.unizh.ch
| Abstract |
|---|
|
|
|---|
Methods: Fluid delivery at steady state conditions, and after vertical displacement of the syringe pump by -50 cm was determined gravimetrically. The Panomat® P-10 infusion pump was evaluated at 4, 10, 20, 50 and 100 µLhr-1, and compared to a conventional syringe pump assembly at 100, 200, 500 and 1000 µLhr-1. Measurements were repeated twice with two different devices of each syringe pump system, and with two syringes. Data are reported as mean ± SD.
Results: Steady state fluid delivery of the Panomat® P-10 infusion pump revealed less than 5% deviation to set flow rate at 10, 20, 50 and 100 µLhr-1, and 12% deviation at 4 µLhr-1. Mean zero-drug delivery time (ZDDT) after lowering the pump by 50 cm at 4 µLhr-1 flow rate was 38.4 ± 7.3 min. At 100 µLhr-1 and with original infusion line ZDDT was almost 20 times shorter when compared to the conventional syringe pump assembly (1.5 ± 0.5 min vs 28.5 ± 5.0 min).
Conclusion: The tested Panomat® P-10 micro-volumetric pump shows an acceptable flow accuracy as well as a low susceptibility to vertical displacement, and is therefore suitable for continuous drug administration at minimal flow rates. The technology used in this pump carries potential implications for a new generation of syringe pumps.
| Introduction |
|---|
|
|
|---|
Because of the known specific problems with conventional syringe pumps at low flow rates we evaluated the Panomat® P-10 micro-volumetric (PMV) infusion pump for intrathecal baclofen administration. The pump was originally designed for subcutaneous long-term application of insulin or iron-chelators. The manufacturer claims precise flow rates down to 1 µLhr-1. However, no data are available about its performance during vertical displacement which in conventional syringe infusion pumps can lead to harmful interruption of drug delivery or to bolus injection.1,4,5 Fluid delivery was therefore assessed after negative vertical displacement at minimal flow rates and compared to a common conventional syringe pump infusion line assembly.
| Methods |
|---|
|
|
|---|
|
Flow rates of 0 µLhr-11000 µLhr-1 and boluses of 0 µL300 µL can be set. Additional features are time lock and limitation of bolus frequency. A continuous flow rate was established by delivering a small bolus every three minutes (three-minute bolus varies according to the preset flow rate).
Conventional syringe pump infusion system
The conventional syringe pump assembly consisted of the Braun Perfusor® compact syringe pump (B. Braun, Melsungen, Germany) combined with a low compliant 50-mL infusion syringe (syringe 5060 mL, Codan Medical Aps, Rodby, DK) and the Panomat® infusion line.
Measurements
For all measurements a gravimetric in vitro set-up was used as published earlier.6 The distal end of the Panomat® or Injectomat® infusion line was connected to an additional stiff infusion line (PE-infusion line, Clinico Medical GmbH, Bad Hersfeld, Germany) immersed with its distal tip by 13 cm in a sampling glass filled with distilled water. A thin layer of oil covered the water surface to avoid fluid evaporation. Assuming a specific weight of distilled water of 1 gmL-1 at room temperature fluid delivery (µLhr-1) into the sampling glass was gravimetrically determined using an electronic balance (AG 204-Delta-Range®, Mettler Toledo, Schwerzenbach, Switzerland; sensitivity 0.0001 g). Data were recorded at a sample rate of one second using a purpose-written software (MCPS V2.6-CAD, Software GmbH, Mönchengladbach, Germany).
First, continuous fluid delivery of the PMV infusion pump was determined at infusion rates of 4, 10, 20, 50, and 100 µLhr-1 each over a one-hour time period after establishing steady state flow. For the measurements the syringe outlet was adjusted to the level of the immersed infusion line tip.
To determine a relative value for the static syringe pump-infusion line compliance, retrograde aspiration volume after lowering the syringe pump by 50 cm was measured while pump was in hold position.
Second, at steady state flow conditions zero-drug delivery time (ZDDT; time period between lowering of the syringe pump and reattainment of initial weight on the balance) was determined after lowering the pump by 50 cm below the level of the immersed infusion line tip. Experiments were performed at infusion rates of 4, 10, 20, 50, and 100 µLhr-1 in the PMV infusion pump, and at 100, 200, 500 and 1000 µLhr-1 in the conventional syringe pump infusion line assembly.
Third, to study the effect of infusion line compliance on ZDDT at very low infusion rates, measurements with the PMV infusion pump were repeated by replacing the original Panomat® infusion line by a soft, more compliant 150 cm infusion line (Injectomat® infusion line; Fresenius AG, Bad Homburg, Germany).
All experiments were performed at 2224°C ambient temperature in duplicates using two individual pumps, and two different syringes with infusion lines of each brand (eight measurements).
Data analysis
Static compliance and ZDDT were compared across the three syringe pump infusion line assemblies by Wilcoxon-Mann-Whitney-Test. Data are presented as mean ± SD. A P value of less than 0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
| Discussion |
|---|
|
|
|---|
Continuous fluid delivery from conventional syringe infusion pumps is known to be a critical issue after vertical displacement of the syringe pump infusion line assembly, and is caused by changes in hydrostatic pressure within the infusion system.1,4 Particularly at low infusion rates, these changes lead to prolonged gravitational siphoning of the infusion solution into the system when the pump device is lowered. Below 1 mLhr-1, drug delivery may be delayed or interrupted for several minutes up to hours after vertical displacement of the infusion pump.5 ZDDT (time with suspended drug delivery) raises with decreasing infusion rates following roughly a 1/X ratio (e.g., halving the infusion rate doubles the ZDDT).5
If the tested conventional syringe pump infusion line assembly, which proved to have the best performance compared to earlier investigations, had been used for the above mentioned intrathecal drug administration at 4 µLhr-1, and assuming that halving the infusion rate doubles the ZDDT, a calculated ZDDT of approximately 13 hr would have resulted.68 Besides interruption of drug administration, it is conceivable that large ZDDTs and retrograde aspiration of blood into a catheter carries the potential risk of accidental catheter occlusion.
Several characteristics may explain the flow accuracy, the outstanding low system compliance and performance after vertical displacement of the PMV infusion pump system:
Instead of the widely used 50-mL syringes in conventional syringe pumps the PMV infusion device is provided with a specially designed 10-mL plastic cartridge. This smaller size has reduced compressible areas, and the avoidance of a compressible rubber seal plunger decreases syringe compliance substantially.6,8 Furthermore, the threaded metallic piston rod, the firm fit of the drive nut into the cartridge body (closed with a screw adapter) and the syringe driver form a stiff and low compliant functional unit.7 Infusion line compliance may further influence pump performance substantially.9 The importance of using the original stiff, low compliant tube extension set with a very small inner diameter was demonstrated by the four-fold increase in ZDDT when it was replaced by the soft, more compliant Injectomat® infusion line (Table II
).9 The design of the tested PMV infusion device results in minimal mechanical gaps, a low overall system compliance and contributes to its optimized functional performance.10
Even though this device may be very advantageous in terms of high flow accuracy and low susceptibility to vertical displacement several points with respect to its potential use in pediatric anesthesia and intensive care medicine must be mentioned: First, in case of accidental occlusion, the devices cut-out pressure is set at 4 bar, resulting in a maximum back-up bolus of 80 µL (manufacturers specifications). This threshold appears excessively high but confirms our measurements of the system compliance (0.02 µLcm-1 H2O). Second, due to the specific pump motor design an appropriate liquid bolus is released stepwise every three minutes (Figure 2
). This fluid delivery strategy is not suitable for inotropic or vasoactive medication since their short half-life (usually one to two minutes) would result in a periodic discontinuity of the desired drug effect. Finally, although the small 10-mL cartridge contributes substantially to the advantages of this specific pump device it would require frequent refilling maneuvers if used for higher fluid delivery rates.
|
In conclusion, the tested PMV infusion pump shows excellent flow accuracy as well as a low susceptibility to negative vertical displacement, and is therefore suitable for drug administration at very low flow rates (e.g., intrathecal as in the case described). The technology used in this pump carries potential implications for a new generation of syringe pumps for its use in pediatric and neonatal intensive care medicine and anesthesia.
| Acknowledgments |
|---|
| Footnotes |
|---|
Revision received August 14, 2002. Accepted for publication April 25, 2002.
| References |
|---|
|
|
|---|
2 Albright AL, Barron WB, Fasick MP, Polinko P, Janosky J. Continuous intrathecal baclofen infusion for spasticity of cerebral origin. JAMA 1993; 270: 24757.[Abstract]
3 Gilmartin R, Bruce D, Storrs BB, et al. Intrathecal baclofen for management of spastic cerebral palsy: multicenter trial. J Child Neurol 2000; 15: 717.
4 Lonnqvist PA. How continuous are continuous drug infusions? (Editorial). Intensive Care Med 2000; 26: 6601.[Medline]
5 Kern H, Kuring A, Redlich U, et al. Downward movement of syringe pumps reduces syringe output. Br J Anaesth 2001; 86: 82831.
6 Weiss M, Hug MI, Neff T, Fischer J. Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anesth 2000; 47: 10315.
7 Neff TA, Fischer JE, Schulz G, Baenziger O, Weiss M. Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensive Care Med 2001; 27: 28791.[Medline]
8 Weiss M, Fischer J, Neff T, Baenziger O. The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 2000; 55: 10948.[Medline]
9 Weiss M, Banziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 2000; 26: 7769.[Medline]
10 Lonnqvist PA, Lofqvist B. Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med 1997; 23: 9981001.[Medline]
This article has been cited by other articles:
![]() |
H. Igarashi, Y. Obata, Y. Nakajima, T. Katoh, K. Morita, and S. Sato Syringe pump displacement alters line internal pressure and flow: [Le deplacement verticalde la pompe a perfusion modifie la pression interne du catheter et le debit] Can J Anesth, August 1, 2005; 52(7): 685 - 691. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |