| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |





* From the Departments of Anesthesiology University of California Los Angeles School of Medicine Charles R. Drew University of Medicine Science Martin Luther King Jr/Drew Medical Center Los Angeles California
The Department Of Clinical Studies New Bolton Center University of Pennsylvania Kennett Square Pennsylvania; and
The Departments Of Anesthesiology Pain Medicine Surgery Medical Pathology University of California-Davis School of Medicine Sacramento California USA.
Dr. J.S. Jahr, Department of Anesthesiology, UC Los Angeles Center for the Health Sciences, Box 951778, Los Angeles, CA 90095, USA. Phone: 310-825-6761; Fax: 310-206-4626; E-mail: jsjahr{at}mednet.ucla.edu
| Abstract |
|---|
|
|
|---|
Methods: In the laboratory, 45 split canine plasma samples were mixed with hemoglobin glutamer-200 (8.13, 16.25, 32.5 gL-1 concentrations), 45 samples were mixed with hemoglobin glutamer-250 (8.13, 16.25, 32.5 gL-1 concentrations), 45 with hemoglobin-raffimer (12.5, 25.0, 50.0 gL-1 concentrations), and measured. Plasma samples without HBOC served as control. Hemoglobin concentration was determined by a laboratory analyzer (Coulter Corporation, Hiafeah, FL, USA) and B-hemoglobin photometer (HemoCue®, Ångelholm, Sweden). Two independent technicians performed blinded sample measurements and randomly tested each sample five times. Results were analyzed according to Bland and Altman analysis.
Results: B-hemoglobin photometer demonstrated high repeatability for all three HBOCs. Repeatability coefficients were 0.37 gL-1 and 0.48 gL-1 for hemoglobin glutamer-200, 0.39 gL-1 and 0.4 gL-1 for hemoglobin glutamer-250 and 1.07 gL-1 and 0.85 gL-1 for hemoglobin-raffimer. An acceptable agreement was found between the B-hemoglobin photometer and the laboratory analyzer for all three HBOCs tested.
Conclusion: The B-hemoglobin photometer accurately determined the concentration of three HBOC solutions dissolved in canine plasma.
| Introduction |
|---|
|
|
|---|
A point-of-care device for measuring blood Hb, the B-hemoglobin photometer (HemoCue®, Ångelholm, Sweden), provides rapid and convenient assessment of hemoglobin concentration. This device uses a chemically pretreated cuvette for the blood sample. The reagents deposited on the inner wall of the cuvette lyse the red cells and convert Hb into azide methemoglobin. The Hb concentration is calculated using spectrophotometric analysis with absorbence peaks at 565 and 880 nm. The manufacturer claims the accuracy of the device to be within ± 3.0 gL-1.1 A comparison between standard laboratory equipment and B-hemoglobin photometer showed a high correlation when using human blood. The HemoCue®, however, has been shown to be more operator-dependent2 than other laboratory hemoglobin analyzers.
The use of hemoglobin-based oxygen carriers under development (HBOC) to diminish allogeneic blood transfusion therapy may require monitoring of the levels of plasma Hb during their administration.3 Therefore, the accuracy of the B-hemoglobin photometer with HBOCs should be validated. This study assessed the accuracy of photometer-based hemoglobin concentration determination using the HemoCue® for three different HBOCs: hemoglobin glutamer-200 (bovine; Oxyglobin®, Biopure Corp, Cambridge, MA, USA), hemoglobin glutamer-250 (bovine; Hemopure®, Biopure Corp, Cambridge, MA, USA), and hemoglobin-raffimer (human; HemolinkTM, Hemosol, Inc., Toronto, Ontario, Canada) in varying concentrations mixed with canine plasma. One of these HBOCs (hemoglobin glutamer-200) is Food and Drug Administration (FDA) approved for veterinary use and indicated for canine anemia; hemoglobin glutamer-250 and hemoglobin-raffimer are currently undergoing or completing FDA phase III trials in the United States, and one, (hemoglobin glutamer-250) is approved for human use in South Africa. The published chemical and physical properties of the three HBOCs are shown in Table I
.
|
| Methods |
|---|
|
|
|---|
In the laboratory, repeated measurements were conducted on samples of canine plasma obtained from one healthy animal mixed with HBOCs in different concentrations. For hemoglobin glutamer-200 and hemoglobin glutamer-250 the concentrations were 8.13, 16.25 and 32.5 gL-1; for hemoglobin-raffimer they were 12.5, 25.0, and 50.0 gL-1. Each dilution was divided into 15 samples (five for each of two technicians, and five for the laboratory analyzer). This resulted in a total of 135 samples, 45 samples for each of the two technicians, and 45 for the laboratory analyzer. Plasma samples containing no HBOC were used as controls. Hemoglobin concentration was determined using both an automated laboratory analyzer (Coulter STCS, Coulter Corporation, Hialeah, FL, USA.) and a bedside B-hemoglobin photometer (HemoCue®). Two experienced, independent laboratory technicians (observer one and two) were blinded to sample number and performed the photometer measurements. Each operator randomly tested each of the 45 samples. The Bland and Altman analysis was used for repeatability estimation, and for measuring agreement between two methods using repeated measurements.7
| Results |
|---|
|
|
|---|
Hemoglobin glutamer-200
The mean difference between two observers was -0.003 ± 0.474 gL-1 and did not depend on hemoglobin concentration (Figure 1
). The mean difference in values between the laboratory analyzer and B-hemoglobin photometer was 0.002 ± 0.116 gL-1. Repeatability coefficients were 0.37 gL-1 for the first observer, 0.48 gL-1 for the second observer, and 0.36 gL-1 for the laboratory analyzer. When comparing values obtained from the laboratory analyzer with those obtained from B-hemoglobin photometer, the 95% confidence interval for the lower limit of agreement was -0.515 to -0.111 gL-1, and for the upper limit of agreement was from 0 to 0.404 gL-1.
|
|
|
| Discussion |
|---|
|
|
|---|
A point of care device, such as the B-hemoglobin photometer appears an appropriate solution for critical care settings and the operating room. Its clinical accuracy and reliability has been confirmed.8,9 Some technical aspects regarding the technology, however, have been emphasized in two reports.9,10 The dependence of the results upon the use of appropriate technique leads to some degree of intra-observer variability.810
The results of hemoglobin concentration measurements with the B-hemoglobin photometer become particularly questionable when concentrations of hemoglobin in plasma are measured, especially as concentrations of the hemoglobin in the tested samples approach the limit of accuracy for the instrument.1 The use of HBOCs makes it clinically important to validate the B-hemoglobin photometer for measurement of plasma hemoglobin. Moreover, some of the HBOCs being under development are heterogeneous hemoglobins or chemically altered hemoglobins; therefore, they have optical properties different from normal human erythrocytic hemoglobin. These possibly can alter the accuracy of photometric measurements.4,11,12
Jaeger et al. compared the HemoCue to a standard hematology analyzer in the clinical setting and demonstrated "sufficient accuracy to support treatment decisions regarding blood transfusions".2 The article generated controversy regarding the use of Bland and Altman analysis,13 which we used to analyze our results. However, the two articles are not comparable, because we tested HBOC's in vitro, while Jaeger et al. tested patients' hemoglobins.
Lardi et al.8 evaluated the B-hemoglobin photometer in the clinical setting, similarly to Jaeger et al. Here again, the results are not comparable to ours, because no HBOC was tested. Rippmann et al.9 studied the B-hemoglobin photometer in vitro and in vivo, compared to a co-oximeter and reported a wider variability, but improved accuracy with multiple averaged evaluations. These results coincide with ours, except the variability appears wider than ours and may be due to the in vitro reconstitution of blood where red cells and fresh frozen plasma were mixed, possibly causing lysis of the red cells.
The study by McNulty et al.10 demonstrated a wide variability of bedside methods of hemoglobin assessment during hemodilution and after transfusion. However, photometry provided reasonably accurate results. This mirrors our own results. In addition we have shown that plasma hemoglobin determinations, in the absence of red blood cells, are accurate. It may be questioned whether our results can be extrapolated to the clinical scenario (dissolved HBOC in the absence of background hemoglobin, compared to the clinical situation of human plasma with red blood cells and HBOC). Obviously, extrapolation to the human clinical setting is impossible. However, veterinary use of HBOCs in canine models suggests this may be valid.11,12 This will require validation in human in vivo models.
In summary, our study shows that, in the clinically relevant range of plasma concentrations of three different HBOCs, the hemoglobin concentrations determined by use of B-hemoglobin Photometer are in agreement with values obtained by use of standard laboratory techniques. Both inter-observer and intra-observer repeatability were acceptable. Based on this in-vitro experiment, we conclude that the B-hemoglobin photometer is an appropriate device for rapid measurement of hemoglobin concentrations when HBOCs are added to plasma.
|
| Acknowledgments |
|---|
Revision received November 21, 2001. Accepted for publication September 26, 2001.
| References |
|---|
|
|
|---|
2
Jaeger M, Ashbury T, Adams M, Duncan P. Perioperative on-site haemoglobin determination: as accurate as laboratory values? Can J Anaesth 1996; 43: 7958.
3
Jahr JS, Lurie F, Xi S, et al. A novel approach to measuring circulating blood volume: the use of a hemoglobin-based oxygen carrier in a rabbit model. Anesth Analg 2001; 92: 60914.
4 Jahr JS, Lurie F, Gosselin R, Lin JS, Wong L, Larkin E. Effects of a hemoglobin-based oxygen carrier (HBOC-201) on coagulation testing. Clin Lab Sci 2000; 13: 2104.[Medline]
5 Kasper SM, Walter M, Grune F, Bischoff A, Kasper SM, Erasmi H, Buzello W. Effects of a hemoglobin-based oxygen carrier (HBOC-201) on hemodynamics and oxygen transport in patients undergoing preoperative hemodilution for elective abdominal aortic surgery. Anesth Analg 1996; 83: 9217.[Abstract]
6 Manning, JE, Katz LM, Brownstein MR, Pearce LB, Gawryl MS, Baker CC. Bovine hemoglobin-based oxygen carrier (HBOC-201) for resuscitation of uncontrolled, exsanguinating liver injury in swine. Shock 2000; 13: 1529.[Medline]
7
Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 13560.
8 Lardi AM, Hirst C, Mortimer AJ, McCollum CN. Evaluation of the HemoCue for measuring intra-operative haemoglobin concentrations: a comparison with the Coulter Max-M. Anaesthesia 1998; 53: 34952.[Medline]
9 Rippmann CE, Nett PC, Popovic D, Seifert B, Pasch T, Spahn DR. Hemocue, an accurate bedside method of hemoglobin measurement? J Clin Monit 1997; 13: 3737.[Medline]
10 McNulty SE, Torjman M, Grodecki W, Marr A, Schieren H. A comparison of four bedside methods of hemoglobin assessment during cardiac surgery. Anesth Analg 1995; 81: 1197202.[Abstract]
11 Jahr JS, Lurie F, Driessen B, et al. Validation of oxygen saturation measurements in a canine model of hemoglobin based oxygen carrier (HBOC) infusion. Clin Lab Sci 2000; 13: 1739.
12 Jahr JS, Driessen B, Lurie F, Tang Z, Louie RF, Kost G. Oxygen saturation measurements in canine blood containing hemoglobin glutamer-200 (bovine): in vitro validation of the Nova CO-oximeter. Vet Clin Path 2001; 30: 3945.
13
Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J. Comparing methods of clinical measurement: reporting standards for Bland and Altman analysis. Anesth Analg 2000; 90: 593602.
This article has been cited by other articles:
![]() |
F. Lurie, B. Driessen, J. S. Jahr, R. Reynoso, and R. A. Gunther Validity of Arterial and Mixed Venous Oxygen Saturation Measurements in a Canine Hemorrhage Model After Resuscitation with Varying Concentrations of Hemoglobin-Based Oxygen Carrier Anesth. Analg., January 1, 2003; 96(1): 46 - 50. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Lurie, J. S. Jahr, and B. Driessen The Novel HemoCue(R) Plasma/Low Hemoglobin System Accurately Measures Small Concentrations of Three Different Hemoglobin-Based Oxygen Carriers in Plasma: Hemoglobin Glutamer-200 (Bovine) (Oxyglobin(R)), Hemoglobin Glutamer-250 (Bovine) (Hemopure(R)), and Hemoglobin-Raffimer (HemolinkTM) Anesth. Analg., October 1, 2002; 95(4): 870 - 873. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |