From the Department of Anesthesiology, University of North Carolina Chapel Hill, North Carolina, USA.
Address correspondence to: Dr. Fred J. Spielman, UNC Department of Anesthesiology, Campus Box 7010, 223 Burnett-Womack Bldg, Chapel Hill, North Carolina 27599-7010, USA. Phone: 919-966-5136; Fax: 919-966-4873; E-mail: fspielman{at}aims.unc.edu
| Abstract |
|---|
|
|
|---|
Clinical features: A 23-yr-old woman underwent a wide local excision of a chest wall melanoma and sentinel node biopsy under general anesthesia. During use of a single-use LMA-UniqueTM, the airway tube became completely separated from the distal mask (backplate). No sequelae resulted from failure of the airway apparatus. The LMA was examined by LMA North America, and the company stated "no definitive explanation can be made to explain the cause for separation of the airway from the backplate."
Conclusion: Although the use of the LMA is associated with rare and minimal complications, each airway should be carefully inspected before its use for loss of integrity resulting from sterilization or for defects in manufacturing.
| Introduction |
|---|
|
|
|---|
|
| Case report |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
Several case reports described the separation,4 transection,5,6 and shattering of LMAs;7 these reports all involved reusable LMAs. Repeated sterilization with heat and chemicals were thought to have caused degradation of the silicone, resulting in brittleness, cracking, and in one case,7 "...the friability of cheddar cheese." On no occasion was it known how many times a LMA had been used.
LMATM North America suggests that the LMA-ClassicTM mask can be reused up to 40 times. Each instance that a reusable airway is considered for use it must be tested for elasticity, and the airway should be discarded if the tube kinks when flexed to 180°.8 The integrity of the cuff ought to be verified by inflating with a volume of air 50% greater than the recommended maximum. There should be a system in place to verify the number of times the LMA has been sterilized.
The LMA was sent to LMATM North America for further inspection and analyses. Upon examination of the LMA-UniqueTM referenced in this case report, LMATM North America noted, "the airway has become completely separated from the backplate. There is evidence of the solvent bond on the area of the airway, which would have inserted into the backplate. The outer diameter of the airway and the inner diameter of the backplate are within specifications and appear to be a good fit when put together. After examination of this LMA-UniqueTM and manufacturing records, no definitive explanation can be made to explain the cause for separation of the airway from the backplate."
Biting on the LMA during the process of device removal might be an important factor in the separation,4 transection,5 and shattering7 of airways. As is recommended, the LMA should be removed with the cuff inflated either after full recovery of airway reflexes or while the patient is deeply anesthetized, with a bite block in place. An advantage to removing the LMA before emergence is that this would diminish the chance of the patient biting and damaging the airway. Two studies9,10 have demonstrated a high incidence of clenched teeth and biting the LMA, 26% and 10%, respectively, when the airway is removed after laryngeal reflexes have returned.
In summary, although the use of the LMA is associated with rare and minimal complications, each airway should be carefully inspected before its use for loss of integrity resulting from sterilization or for defects in manufacturing.
Revision received August 9, 2002. Accepted for publication May 17, 2002.
| References |
|---|
|
|
|---|
2 Brimacombe J. Analysis of 1500 laryngeal mask uses by one anaesthetist in adults undergoing routine anaesthesia. Anaesthesia 1996; 51: 7680.[Medline]
3 Verghese C, Brimacombe JR. Survey of laryngeal mask airway usage in 11,910 patients: safety and efficacy for conventional and nonconventional usage. Anesth Analg 1996; 82: 12933.[Abstract]
4 Khoo ST. The laryngeal mask airwayan unusual complication (Letter). Anaesth Intensive Care 1993; 21: 24950.
5 Kramer-Kilper OT. Removal of laryngeal mask airway during light anaesthesia (Letter). Anaesthesia 1992; 47: 816.
6 Crawford M, Davidson G. A problem with a laryngeal mask airway (Letter). Anaesthesia 1992; 47: 76.
7 Squires SJ. Fragmented laryngeal mask airway (Letter). Anaesthesia 1992; 47: 274.
8 Asai T, Vaughan RS. Misuse of the laryngeal mask airway (Editorial). Anaesthesia 1994; 49: 4679.[Medline]
9 Gataure PS, Latto IP, Rust S. Complications associated with removal of the laryngeal mask airway: a comparison of removal in deeply anaesthetised versus awake patients. Can J Anaesth 1995; 42: 11136.
10 Reddy SVG, Win N. Brain laryngeal mask - study in 50 spontaneously breathing patients. Singapore Med J 1990; 31: 33840.[Medline]
This article has been cited by other articles:
![]() |
R. A. Jee and G. L. van Hasselt Manufacturer's response to a defective cuff in a disposable laryngeal mask airway Can J Anesth, August 1, 2007; 54(8): 681 - 682. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||