CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Abstract Freely available
Right arrow Résumé de cet Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Spielman, F. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Spielman, F. J.
Canadian Journal of Anesthesia 49:990-992 (2002)
© Canadian Anesthesiologists' Society, 2002

Cardiothoracic Anesthesia, Respiration and Airway

Complete separation of the tube from the mask during removal of a disposable laryngeal mask airway

[La séparation complète du tube du masque pendant le retrait d’un masque laryngé jetable]

Fred J. Spielman, MD

From the Department of Anesthesiology, University of North Carolina Chapel Hill, North Carolina, USA.

Address correspondence to: Dr. Fred J. Spielman, UNC Department of Anesthesiology, Campus Box 7010, 223 Burnett-Womack Bldg, Chapel Hill, North Carolina 27599-7010, USA. Phone: 919-966-5136; Fax: 919-966-4873; E-mail: fspielman{at}aims.unc.edu


    Abstract
 TOP
 Abstract
 Introduction
 Case report
 Discussion
 References
 
Purpose: To describe a complication of the disposable laryngeal mask airway (LMA).

Clinical features: A 23-yr-old woman underwent a wide local excision of a chest wall melanoma and sentinel node biopsy under general anesthesia. During use of a single-use LMA-UniqueTM, the airway tube became completely separated from the distal mask (backplate). No sequelae resulted from failure of the airway apparatus. The LMA was examined by LMA North America, and the company stated "no definitive explanation can be made to explain the cause for separation of the airway from the backplate."

Conclusion: Although the use of the LMA is associated with rare and minimal complications, each airway should be carefully inspected before its use for loss of integrity resulting from sterilization or for defects in manufacturing.


    Introduction
 TOP
 Abstract
 Introduction
 Case report
 Discussion
 References
 
SINCE Brain’s1 publication about the laryngeal mask airway (LMA; LMATM North America, San Diego, CA, USA) in 1983, the device has made a significant contribution to airway control. The LMA is popular because it is easy to position, provides a secure airway for spontaneously breathing patients, and affords effective assisted ventilation in elective and emergency situations without requiring endotracheal intubation or visualization of the glottis. Complications associated with the LMA have been reported to occur at insertion, during anesthesia, and upon emergence; however, the difficulties are rare and usually minor. This report describes the occurrence of difficulties with an unused LMA-UniqueTM (non-reusable) where the airway tube completely separated from the distal mask (FigureGo).



View larger version (112K):
[in this window]
[in a new window]
 
FIGURE The airway tube completely separated from the mask (backplate).

 

    Case report
 TOP
 Abstract
 Introduction
 Case report
 Discussion
 References
 
A 23-yr-old woman weighing 58 kg underwent a wide local excision of a chest wall melanoma and sentinel node biopsy under general anesthesia. After iv administration of midazolam 2 mg, fentanyl 50 µg, lidocaine 60 mg, and propofol 150 mg, a size 3 disposable LMA was inserted without difficulty. A soft bite block was used. General anesthesia was maintained with sevoflurane, air, and oxygen, while the patient breathed spontaneously. An attempt was made to remove the LMA at the conclusion of the 93-min operation, when the patient followed commands. No biting was noted on the airway device. A moderate degree of traction was applied to the airway device. The proximal tube of the LMA slid out of the mouth, leaving the distal mask inside the pharynx still attached to the pilot balloon. The remaining portion of the LMA was easily removed by grasping the pilot balloon tubing. At no time during the extubation did the patient’s oxygen saturation decrease below 98%. The patient was discharged home two hours later without apparent complications. Assessment of the LMA showed that the distal mask portion had obviously separated from the proximal tube where they were joined (FigureGo).


    Discussion
 TOP
 Abstract
 Introduction
 Case report
 Discussion
 References
 
Complications are infrequent during use of the LMA in the operating room. The most common problems encountered include inability to position the LMA correctly, coughing and gagging during placement and removal, laryngospasm, and postoperative sore throat. Mechanical problems are rare. Brimacombe’s2 analysis of 1500 LMA uses revealed that on two occasions (0.13%) the LMA cuff slowly deflated during the operation secondary to a small leak. Verghese and Brimacombe3 found that only 44 critical incidents were documented as related to the use of a LMA in 11,910 anesthetics (0.37%), and only 18 (0.15%) of these were related to the airway. None of the incidents concerned structural issues associated with the LMA.

Several case reports described the separation,4 transection,5,6 and shattering of LMAs;7 these reports all involved reusable LMAs. Repeated sterilization with heat and chemicals were thought to have caused degradation of the silicone, resulting in brittleness, cracking, and in one case,7 "...the friability of cheddar cheese." On no occasion was it known how many times a LMA had been used.

LMATM North America suggests that the LMA-ClassicTM mask can be reused up to 40 times. Each instance that a reusable airway is considered for use it must be tested for elasticity, and the airway should be discarded if the tube kinks when flexed to 180°.8 The integrity of the cuff ought to be verified by inflating with a volume of air 50% greater than the recommended maximum. There should be a system in place to verify the number of times the LMA has been sterilized.

The LMA was sent to LMATM North America for further inspection and analyses. Upon examination of the LMA-UniqueTM referenced in this case report, LMATM North America noted, "the airway has become completely separated from the backplate. There is evidence of the solvent bond on the area of the airway, which would have inserted into the backplate. The outer diameter of the airway and the inner diameter of the backplate are within specifications and appear to be a good fit when put together. After examination of this LMA-UniqueTM and manufacturing records, no definitive explanation can be made to explain the cause for separation of the airway from the backplate."

Biting on the LMA during the process of device removal might be an important factor in the separation,4 transection,5 and shattering7 of airways. As is recommended, the LMA should be removed with the cuff inflated either after full recovery of airway reflexes or while the patient is deeply anesthetized, with a bite block in place. An advantage to removing the LMA before emergence is that this would diminish the chance of the patient biting and damaging the airway. Two studies9,10 have demonstrated a high incidence of clenched teeth and biting the LMA, 26% and 10%, respectively, when the airway is removed after laryngeal reflexes have returned.

In summary, although the use of the LMA is associated with rare and minimal complications, each airway should be carefully inspected before its use for loss of integrity resulting from sterilization or for defects in manufacturing.

Revision received August 9, 2002. Accepted for publication May 17, 2002.


    References
 TOP
 Abstract
 Introduction
 Case report
 Discussion
 References
 
1 Brain AIJ. The laryngeal masks—a new concept in airway management. Br J Anaesth 1983; 55: 801–5.[Abstract/Free Full Text]

2 Brimacombe J. Analysis of 1500 laryngeal mask uses by one anaesthetist in adults undergoing routine anaesthesia. Anaesthesia 1996; 51: 76–80.[Medline]

3 Verghese C, Brimacombe JR. Survey of laryngeal mask airway usage in 11,910 patients: safety and efficacy for conventional and nonconventional usage. Anesth Analg 1996; 82: 129–33.[Abstract]

4 Khoo ST. The laryngeal mask airway—an unusual complication (Letter). Anaesth Intensive Care 1993; 21: 249–50.

5 Kramer-Kilper OT. Removal of laryngeal mask airway during light anaesthesia (Letter). Anaesthesia 1992; 47: 816.

6 Crawford M, Davidson G. A problem with a laryngeal mask airway (Letter). Anaesthesia 1992; 47: 76.

7 Squires SJ. Fragmented laryngeal mask airway (Letter). Anaesthesia 1992; 47: 274.

8 Asai T, Vaughan RS. Misuse of the laryngeal mask airway (Editorial). Anaesthesia 1994; 49: 467–9.[Medline]

9 Gataure PS, Latto IP, Rust S. Complications associated with removal of the laryngeal mask airway: a comparison of removal in deeply anaesthetised versus awake patients. Can J Anaesth 1995; 42: 1113–6.[Abstract/Free Full Text]

10 Reddy SVG, Win N. Brain laryngeal mask - study in 50 spontaneously breathing patients. Singapore Med J 1990; 31: 338–40.[Medline]




This article has been cited by other articles:


Home page
Canadian J. AnesthesiaHome page
R. A. Jee and G. L. van Hasselt
Manufacturer's response to a defective cuff in a disposable laryngeal mask airway
Can J Anesth, August 1, 2007; 54(8): 681 - 682.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Résumé de cet Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Spielman, F. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Spielman, F. J.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS