| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |

* From the Department of Psychology, University of Manitoba, Winnipeg, Manitoba, and
the Department of Anesthesiology, Dalhousie University, Halifax, Nova Scotia, Canada
Address correspondence to: Dr. Todd Mondor, Department of Psychology, P429 Duff Roblin Bldg., University of Manitoba, Winnipeg, Manitoba R3T 2N2, Canada. Phone: 204-474-7837; Fax: 204-474-7599; E-mail: todd_mondor{at}umanitoba.ca
| Abstract |
|---|
|
|
|---|
Methods: Undergraduate students, who were naïve with respect to the clinical situation associated with the alarms, judged perceived urgency of each alarm on a ten-point scale.
Results: The perceived urgency of the alarms was not consistent with the actual urgency of the clinical situation that triggers it. In addition, those alarms indicating patient condition were generally perceived as less urgent than those alarms indicating the operation of equipment. Of particular interest were three sets of alarms designed by equipment manufacturers to indicate specific priorities for action. Listeners did not perceive any differences in the urgency of the information only, medium and high priority alarms of two of the monitors with all judged as low to moderate in urgency. In contrast, the high priority alarm of the third monitor was judged as significantly more urgent than its low and medium urgency counterparts.
Conclusion: The alarms currently in use do not convey the intended sense of urgency to naïve listeners, and this holds even for two sets of alarms designed specifically by manufacturers to convey different levels of urgency.
| Introduction |
|---|
|
|
|---|
Some of these changes in alarm design have occurred in response to recently adopted international standards for auditory warning alarm design.9 Although permitting some latitude in alarm design, the ISO recommendations (Tables I and II![]()
) set out clear limits for the design of medium and high priority alarms, including such factors as the dominant frequency of an alarm, the number of individual pulses that may comprise a burst, and the repetition rate of both the pulses within a burst and of the bursts themselves.
|
|
One of the principal motivations in establishing these recommendations for the design of auditory warning signals appears to have been "to have a pattern which is instantly recognizable to the trained respondent, but which will not usually evince anxiety in others".9 Thus, the guidelines established by ISO are intended to facilitate the generation of alarms that are clear and easy to interpret but that are not so strident as to upset patients and their families. The recommendations also speak to another important element of alarm design; namely, the importance of tailoring the acoustic properties of each alarm to the urgency of the triggering situation. In particular, the ISO standards include a suggestion that the time between pulses within a burst and the most prominent frequency component of a pulse be modulated to exert some control over the sense of urgency induced in a listener by the alarm. This recommendation is consistent with research indicating that basic auditory features such as pitch, loudness, and repetition rate may have a dramatic influence on the perception of the urgency of an alarm.3,10,11 Thus, a sound with a high pitch or a rapid repetition rate will be perceived as more urgent than a sound with a lower pitch or repetition rate. An auditory warning alarm designed with these principles in mind could reduce the need for interpretation of urgency and, thereby, reduce the possibility of a misclassification leading to an inappropriate response.
The present study was designed primarily to evaluate the perceived urgency of a set of patient and equipment auditory warning alarms that are common in the hospital operating room. In addition, however, we were interested in examining the perceived urgency of three sets of alarms specifically designed by equipment manufacturers to represent different levels of clinical priority. To this end, specific comparisons of perceived urgency were conducted of the alarms used in the Datex AS/3 monitor (information only, medium and high priority alarms), the Hewlett-Packard model 66S monitor (low, medium and high priority), and the North American Dräger Vitalert 3200 (medium and high priority).
| Methods |
|---|
|
|
|---|
2. Materials
COMPUTER AND SOUND SYSTEM
A 486/150 personal computer running the Micro-Experimental Laboratory12 was used to present sounds and to record responses. Sounds were presented over Sennheiser HD 250 linear headphones.
ALARMS
Thirteen alarms were recorded at the IWK-Grace Health Centre and the Queen Elizabeth II Health Sciences Centre in Halifax, Nova Scotia, Canada, on a Dell P133ST laptop computer at a sampling rate of 32,000 Hz. Background noise was removed from the recordings using the filtering functions of CoolEdit 96.13 All alarms were equated in subjective intensity and presented to listeners in the experiment at a comfortable volume of approximately 70 decibels sound pressure level. The acoustic characteristics of each of the 13 alarms used in the study are described in Table III
. This summary is based on a frequency analysis performed using CoolEdit 96.
|
| Results |
|---|
|
|
|---|
|
|
| Planned comparisons of alarm sets designed by equipment manufacturers |
|---|
|
|
|---|
North American Dräger Vitalert 3200
The North American Dräger Vitalert 3200 includes two alarms, one intended to indicate a medium priority situation and the other intended to indicate a high priority situation. Participants judged these alarms as essentially equivalent in urgency (3.138, and 3.127, respectively). In addition, and as was the case for the Datex alarms, neither alarm was perceived as indicating anything more than quite a low level of urgency.
Hewlett-Packard model 66S monitor
Whereas the alarms designed by Datex and by North American Dräger did not convey the intended sense of urgency to our subjects, the set of three alarms designed by Hewlett-Packard for the model 66S monitor did induce, to at least some degree, differences in perceived urgency. Specifically, although listeners did not differentiate between the low priority and medium priority alarms (mean ratings of 4.896 and 4.522 respectively), they did perceive the high priority alarm as possessing a significantly higher urgency (7.127).
| Discussion |
|---|
|
|
|---|
Previous research has shown that sound properties such as frequency composition, repetition rate, amplitude, and harmonic relation of the frequency components can significantly influence the interpretation of the urgency of an auditory alarm (e.g., 3, 11). An examination of the fundamental frequencies of the four alarms judged as most urgent indicates that they were based on the four highest fundamental frequencies in the set of 13 alarms [Bair (2600 Hz), Ritchie (33003600 Hz), Hotline (3360 Hz), and HP-high priority (2900 Hz)]. In addition, although the Ritchie whistle is continuous, the other three alarms all have quite a rapid repetition rate (Table III
). The Bair alarm, in particular, which was judged as most urgent has both a high frequency combined with the most rapid repetition rate of all 13 alarms. In contrast the alarm judged as least urgent (Datex AS/3 monitor: information/BP ready") consists of a range of harmonically-related frequency components (F0 = 250 Hz with even harmonics F2 to F12) with a slow repetition rate (one second between pulses). It is interesting to note that for those alarms that include multiple frequency components these are always harmonically-related. Edworthy et al. have shown that complex sounds with inharmonic relations are often judged as higher in perceived urgency than sounds that include only harmonic relations.3 A harmonic structure has undoubtedly been chosen by manufacturers in an attempt to generate alarms that are pleasing to the ear, perhaps in hopes of encouraging physicians to use them. Table III
includes a description of the frequency composition and repetition characteristics of each of the 13 alarms.
The importance of basic acoustic characteristics in determining the perceived urgency of an alarm was apparent for the alarm sets designed by manufacturers. For example, the urgency of the three alarms of the Datex monitor AS/3 was judged as essentially equivalent in perceived urgency even though they are intended to convey distinct levels of urgency. Although the frequency composition and the number of pulses do differ somewhat across the three alarms for all of the alarms, the fundamental frequency on which a pulse is based is never higher than 990 Hz. This is important because fundamental frequency carries a significant weight in determining the overall pitch of a sound14 with higher frequencies associated with higher levels of urgency.3 Interestingly, although the repetition rate (operationally defined as the time between successive pulses) for the medium and high priority alarms is quite fast, this did not influence urgency ratings. Thus, pitch appeared to dominate in determining the perception of urgency for these three alarms. Finally, these alarms appear to have been designed to convey urgency information primarily through the number of pulses incorporated within each alarm (one burst of two pulses, one burst of three pulses, and two bursts of five pulses for the information, medium priority, and high priority alarms respectively). This would appear to be a poor choice as the attending physician must wait until the end of the alarm to interpret its urgency.
The Vitalert 3200 includes medium and high priority alarms, both of which consist of three pulses that flow smoothly from one to the next. Thus, the individual pulses are not perceived as discrete but as part of a continuous sound. Although the fundamental frequency of the first two pulses of the high priority alarm (1275 Hz) is higher than that for the medium priority alarm (850 Hz) this is quite a modest difference, and both fall in a low range. Moreover, the slow onset and offset times of the pulses within each alarm (approximately 60 msec ramps for the medium priority alarm and approximately 100 msec ramps for the high priority alarm) would appear also to create the impression of low urgency.3
In contrast with the other two sets of alarms, the three alarms used in the HP monitor did engender some difference in the perception of urgency with the high priority alarm judged as significantly more urgent than the other two. The low and medium priority alarms, which were not perceived as different in urgency, both consist of fairly long duration pulses (650 msec and 750 msec respectively) with long silent periods between pulses (1500 msec for both alarms). In addition to this slow repetition rate, both alarms are based on a very low fundamental frequency of 500 Hz. In contrast, the high priority alarm is composed of a single frequency of 2900 Hz and although the pulses are 850 msec in duration the time between successive pulses is quite brief at 150 msec. Thus, the high priority alarm is both higher in pitch and has a faster repetition rate than the other two alarms. The design of this high priority alarm is, of course, consistent with the known influence of pitch and repetition rate on perceived urgency.
Participants in the study were naïve with respect to the relation between individual alarms and the clinical situations that trigger them. It is unlikely, then, that the urgency ratings were influenced by any prior familiarity with alarms because those used in the study are quite different from the everyday alarms people are familiar with (sirens, fire alarms, buzzers of various kinds). Thus, it seems reasonable to assume that the urgency ratings we obtained are founded primarily on an interpretation of the physical characteristics of the alarms themselves. However, quite aside from basic acoustic composition, interpretation of the situational urgency indicated by an alarm may be determined by the accumulated knowledge and experience of the physician. For example, in pulse oximetry increasing urgency is represented by a reduction in pitch. One might argue that elementary physical features such as pitch and repetition rate are of little consequence when the perceiver is familiar with the relation between individual alarms and the urgency of the triggering situation. As seductive as this argument may be, it ignores the fact that alarms are likely to sound when the attention of the physician is engaged elsewhere. Literally hundreds of studies have shown that the accuracy of perception and identification of an auditory (or visual) stimulus may be significantly influenced, either positively or negatively, by the direction of attention.1517 More importantly, those elements of an event that depend on complex cognitive processing or interpretation are most severely compromised when insufficient attention is available.15 For example, when attention is directed to one verbal message, a change in the voice speaking a second message will likely be noticed but the semantic content of the unattended message will likely be missed.18,19 Given the conditions in which alarms are likely to sound, such research indicates that the physical properties of the alarm are much more likely to be accurately perceived than are those attributes that require deliberate analysis and interpretation. For this reason, auditory alarms that must be interpreted to support an accurate determination of urgency, such as most of those used in the current study, would appear to be ill-suited to the clinical situation (see also Block,20 who proposed a set of melodic alarms that depend entirely on interpretation). It is critical that in designing alarms as much information as possible be included within the elementary physical features of the sound itself. Such thoughtful design would facilitate rapid differentiation of urgency and, thereby, permit physicians to establish priorities for rapid action.
In summary, it is important to note that auditory alarms can serve an important role in alerting physicians to the presence of a potential emergency situation. Indeed, because visual alarms must be fixated to be perceived and interpreted, auditory alarms are potentially much more useful in capturing attention in many settings including the operating room.21 Ideally, the physical composition of each alarm should be specifically designed to convey a perception of urgency that matches, as closely as possible, the actual urgency of the triggering situation. In this way, the cognitive processing required for interpretation of the urgency of a situation signalled by an alarm may be reduced, thereby increasing the speed and appropriateness of the operators response.7,8 In contrast with this ideal design, our examination indicates that the alarms currently in use do not convey the intended sense of urgency to naïve listeners. More critically, this failure applies also to two of the three sets of alarms designed specifically by manufacturers, with the Datex AS/3 monitor information only, medium priority and high priority alarms and the North American Dräger Vitalert 3200 medium and high priority alarms all judged as low to moderate in urgency with no significant differences between them. Although the Hewlett-Packard model 66S monitor high priority alarm was judged as significantly higher in urgency than its low and medium priority alarms there was no difference in urgency of the latter two. Clearly, manufacturers would do well to engage in more thoughtful alarm design.
| Footnotes |
|---|
Revision received November 4, 2002. Accepted for publication July 12, 2002.
| References |
|---|
|
|
|---|
2 Finley GA, Cohen AJ. Perceived urgency and the anaesthetist: responses to common operating room monitor alarms. Can J Anaesth 1991; 38: 95864.
3 Edworthy J, Loxley S, Dennis I. Improving auditory warning design: relationship between warning sound parameters and perceived urgency. Human Factors 1991; 33: 20531.[Medline]
4 Sorkin, RD. Why are people turning off our alarms? J Acoust Soc Am 1988; 84: 11078.
5 Stanford LM, McIntyre JWR, Nelson TM, Hogan JT. Affective responses to commercial and experimental auditory alarm signals for anaesthesia delivery and physiological monitoring equipment. Int J Clin Monit Comput 1988; 5: 1118.[Medline]
6 Weinger MB, Englund CE. Ergonomic and human factors affecting anesthetic vigilance and monitoring performance in the operating room environment. Anesthesiology 1990; 73: 9951021.[Medline]
7 Burt JL, Bartolome DS, Burdette DW, Comstock JR Jr. A psychophysiological evaluation of the perceived urgency of auditory warning signals. Ergonomics 1995; 38: 232740.[Medline]
8 Sorkin RD, Kantowitz BH, Kantowitz SC. Likelihood alarm displays. Human Factors 1988; 30: 44559.
9 CAN/CSA-ISO 9703.2-97 Anaesthesia and respiratory care alarm signals Part 2: auditory alarms signals. Can Standards Assoc., 1997.
10 Patterson RD. CAA Paper 82017. Guidelines for auditory warning systems on civil aircraft. London: UK, Civil Aviation Authority, 1982.
11 Patterson RD. Auditory warning sounds in the work environment. Phil Trans R Soc Lond 1990; 327: 48592.[Medline]
12 Schneider W. Micro experimental laboratory: an integrated system for IBM PC compatibles. Behav Res Methods Instrum Comput 1988; 20: 20617.
13 Syntrillium Software Corp. (1996). Cool Edit 96. Scottsdale, AZ. http://www.syntrillium.com/
14 Moore BCJ. An Introduction to the Psychology of Hearing. London: Academic Press; 1991.
15 Broadbent DE. Perception and Communication. Oxford: Oxford University Press; 1958.
16 Pashler HE. The Psychology of Attention. Cambridge, MA: MIT Press; 1998.
17 Norman DO. The Design of Everyday Things. New York: Basic Books, 1988.
18 Moray N. A data base for theories of selective listening. In: Rabbitt PMA, Dornic S. (Eds.). New York: Academic Press; 1974: 11935.
19 Cowan N. Attention and Memory. An Integrated Framework. Oxford: Oxford University Press, 1995.
20 Block FE Jr. Evaluation of users abilities to recognize musical alarm tones. J Clin Monit 1992; 8: 2832.[Medline]
21 Morris RW, Montano SR. Response times to visual and auditory alarms during anaesthesia. Anaesth Intensive Care 1996; 24: 6824.[Medline]
This article has been cited by other articles:
![]() |
D. M. Korniewicz, T. Clark, and Y. David A National Online Survey on the Effectiveness of Clinical Alarms Am. J. Crit. Care., January 1, 2008; 17(1): 36 - 41. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |