CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Abstract Freely available
Right arrow Résumé de cet Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hemmerling, T. M.
Right arrow Articles by Donati, F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hemmerling, T. M.
Right arrow Articles by Donati, F.
Canadian Journal of Anesthesia 51:116-121 (2004)
© Canadian Anesthesiologists' Society, 2004

General Anesthesia

Comparison of phonomyography with balloon pressure mechanomyography to measure contractile force at the corrugator supercilii muscle

[Comparaison de la phonomyographie et de la mécanomyographie avec ballonnet sous pression pour mesurer la force contractile du muscle sourcilier]

Thomas M. Hemmerling, MD DEAA, Guillaume Michaud, Denis Babin, MSc, Guillaume Trager, DESS and François Donati, MD PhD FRCPC

From the Neuromuscular Research Group, Department of Anesthesiology, Centre hospitalier de l’université de Montréal (CHUM), Hôtel-Dieu, Université de Montréal, Montréal, Québec, Canada.

Address correspondence to: Dr. T.M. Hemmerling, Department of Anesthesiology, Centre hospitalier de l’université de Montréal (CHUM), Hôtel-Dieu, 3840 rue St. Urbain, Montréal, Québec H2W 1T8, Canada. Phone: 514-890-8000, ext. 14570; Fax: 514-412-7222; E-mail: thomashemmerling{at}hotmail.com


    Abstract
 TOP
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
Purpose: Phonomyography is based on the creation of low frequency sounds during muscle contraction, which can be recorded and used for neuromuscular monitoring. In this study, balloon pressure mechanomyography, a novel method to measure the force of contraction via pressure changes in an air-filled balloon, was compared with phonomyography to determine neuromuscular blockade at the corrugator supercilii muscle.

Method: After approval of the Ethics Committee and informed consent, 15 patients were studied. A small condenser microphone was taped to the area just above the eyebrow for phonomyography; an air-filled balloon was taped to the area just above the opposite eyebrow. After induction of anesthesia using remifentanil and propofol, a laryngeal mask airway was inserted without the aid of neuromuscular blocking agents. The facial nerve was stimulated supramaximally with single-twitch stimulation (0.1 Hz) using superficial electrodes placed on both temporal areas for onset and train-of-four stimulation every 12 sec during offset of neuromuscular blockade produced by mivacurium 0.1 mg•kg-1. Onset and recovery measured by the two methods were compared using the t test and agreement between phonomyography and balloon pressure mechanomyography was examined using the Bland-Altman method.

Results: Onset, peak effect, and time to reach 25%, 75%, and 90% of control twitch response for phonomyography vs balloon pressure method were 83 ± 16 sec vs 81 ± 15 sec, 80 ± 15% vs 82 ± 17%, 7.7 ± 2.3 min vs 7.5 ± 2.4 min, 9.9 ± 4.1 min vs 10.5 ± 4 min, and 12.6 ± 4.3 min vs 13.1 ± 4.5 min respectively without being significantly different. Mean bias was 1% with limits of agreement of -9 and +9% of twitch height (T1).

Conclusion: We applied a balloon pressure method to measure the force at the corrugator supercilii. Phonomyography at the corrugator supercilii shows good agreement with this modified version of mechanomyography.


    Introduction
 TOP
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
MECHANOMYOGRAPHY has long been regarded as the gold standard of neuromuscular monitoring because it measures the actual force created by muscle contraction. However, it has its limitations, notably unstable signals and difficult set-up procedures, which necessitate special arm boards. Furthermore, it cannot be used easily at several muscles of interest, such as the corrugator supercilii, which reflects neuromuscular blockade at more profound muscles, such as the larynx, better than the adductor pollicis.1 Validation of other methods of neuromuscular monitoring at the corrugator supercilii muscle have been impaired by the fact that measuring the force of contraction of this small muscle of the eyebrow is not readily possible with the same force transducers used for mechanomyography at the adductor pollicis muscle. Therefore, we have developed a mechanomyographic method similar to the method for measuring the force of contraction of the diaphragm or the adductor laryngeal muscles.2

Recently, phonomyography has been described as a new method of neuromuscular monitoring.3 The method is based on the fact that muscle contraction evokes sounds of low frequency, which can be recorded using special microphones.4,5 Initially, this method was called acoustic myography6,7 and used bulky air-chamber microphones, which were difficult to apply and were unable to detect frequencies lower than 20 Hz. Results from early studies were discrepant for this method’s performance and agreement with established methods of neuromuscular monitoring. In recent studies,3,8,9 we used small condenser microphones to detect lower frequencies better. We determined signal characteristics and frequency response of neuromuscular blockade at the corrugator supercilii and could show that peak frequencies occur at 3 to 5 Hz.3 We also stressed the importance of placement of the microphone in relation to the muscle to optimize signal recording. This study uses phonomyography to measure neuromuscular blockade at the corrugator supercilii muscle and determines its agreement with balloon pressure mechanomyography.


    Patients and methods
 TOP
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
The study was approved by the local Ethics Committee. After obtaining informed consent, 15 patients undergoing general surgery were included. Patients with co-existing neuromuscular disease or patients on medication known to interact with neuromuscular transmission were excluded.

After arrival in the operating theatre, routine monitors (non-invasive blood pressure cuff, pulse oximeter, five-lead electrocardiography) were applied. Anesthesia was induced with iv remifentanil 0.25 to 0.5 µg•kg-1•min-1 followed by iv propofol 2 to 3 mg•kg-1 two minutes later. After loss of consciousness and ventilation via facemask for two minutes with 100% oxygen, a laryngeal mask airway (size 4 for women, size 5 for men, Laryngeal Mask Company, Oxon, UK) was inserted and controlled ventilation commenced with minute ventilation set to maintain an end-tidal PETCO2 of 25 to 40 mmHg. Anesthesia was maintained with a 1 to 1.5 minimum alveolar concentration of sevoflurane in a mixture of 30% oxygen in air. Analgesia was provided by remifentanil 0.05 to 0.25 µg•kg-1•min-1 throughout surgery.

The contractile force of the corrugator supercilii muscle was measured using a small air-filled balloon, filled with 3 mL of air, connected to a pressure transducer (Figure 1GoGo). The balloon was glued to the skin to make sure that all the sides were completely attached to the skin in order to avoid detachment of the balloon during muscle contraction. To restrict upward movements of the balloon away from the skin, a small piece of plastic was glued on top of the balloon and tightly taped to the ipsilateral temporal forehead and contralateral frontal area. The pressure signal was amplified using an AC/DC amplifier (Model 7P122, Grass Instruments, Astra-Med Inc., West Warwick, RI, USA) and its amplitude was measured (Figure 2Go). Changes of intra-balloon pressure reflected muscle contractile force as described for the larynx.2



View larger version (107K):
[in this window]
[in a new window]
 
FIGURE 1A Photograph of the balloon. Pressure changes in the air-filled balloon were used to measure the force created by the contraction of the corrugator supercilii muscle. The 2.5 x 2.5 cm balloon with flattened sides can be firmly glued to the skin to restrict upward movement of the balloon away from the skin.

 


View larger version (96K):
[in this window]
[in a new window]
 
FIGURE 1B Photograph of the measurement setup. The balloon is attached to the left corrugator supercilii muscle with a small piece of plastic to restrict movement away from the skin. This prevents underestimation of the actual force and serves as a form of pre-load.

 


View larger version (27K):
[in this window]
[in a new window]
 
FIGURE 2 Phonomyographic and balloon pressure mechanomyographic signals after train-of-four stimulation. Arrows indicate amplitudes measured.

 
The acoustic signals produced by the contraction of the opposite corrugator supercilii muscle were recorded as described before.3 The microphone signal was amplified and band pass filtered between 0.5 Hz and 1000 Hz using an AC/DC amplifier (Model 7P122, Grass Instruments, Astra-Med Inc., West Warwick, RI, USA). The signals were continuously sampled at 100 Hz using the Polyview® software package (Astra-Med, Inc., West Warwick, RI, USA), digitized, and stored on a portable microcomputer. The single-twitch phonomyographic signal was measured peak-to-peak (Figure 2Go).

Neuromuscular block was measured by mechanomyography and phonomyography simultaneously (Figure 1bGo) with a time shift of two seconds between the two sides, which was necessary to avoid measurement of contraction of the contralateral muscle. In all patients, the facial nerve was stimulated at the temporal area after searching for the best possible stimulation site. Facial nerve stimulation in this area creates contraction of the orbicularis oculi muscle and the corrugator supercilii muscle; therefore, displacement of the stimulating electrodes to achieve optimal contraction of the corrugator supercilii muscle with minimal contraction of the orbicularis oculi muscle is important. Usually, this is done by shifting the electrodes cranially. Once the best stimulation site was found, a current of 20 to 30 mA was set to achieve supramaximal stimulation, which was determined using single-twitch stimulation at 0.1 Hz. Stimulation was performed via surface electrodes using a constant current stimulator (Innervator®, Fisher and Paykel Healthcare, Auckland, New Zealand) that generated single-twitch square pulses of 0.2 msec with a current intensity between 0 and 70 mA [train-of-four (TOF) stimulation every 12 sec for offset of neuromuscular blockade, with a time shift of two seconds between the two sides]. After at least five minutes of supramaximal stimulation and stable baselines for both recordings, iv mivacurium 0.1 mg•kg-1 was injected within five seconds into a fast flowing solution of Ringer’s lactate. Onset, maximal effect, and offset of neuromuscular blockade after mivacurium were determined.

The first twitch response was used to analyze the time to reach maximum decrease of twitch response (onset time), and the times to reach 25% (T 25%), 75% (T 75%), and 90% (T 90%) of control twitch response. The maximum effect was defined as the maximal decrease of the twitch response. Time to reach a TOF ratio of 0.5, 0.7, 0.8 and 0.9 was calculated for balloon pressure mechanomyography and phonomyography simultaneously.

Sample size was calculated to detect an estimated difference of 20% of mean onset time and T 25% between the two methods with a type I error rate of 5% and a power of 95%. The mean difference of all signals determined using both methods (mechanomyography minus phonomyography) and the limits of agreement (LOA) between the two methods were analyzed using the Bland-Altman method.10 Correlation between the two methods was calculated using Pearson’s test. Data were compared between phonomyography and balloon pressure mechanomyography using the t test. A P-value < 0.05 was considered statistically significant. Data are presented as mean ± standard deviation.


    Results
 TOP
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
In all patients, pharmacodynamic results with both methods could be obtained. Recordings of signals were continued until TOF ratios were greater than 0.9 in all patients. Typical phonomyographic and balloon pressure mechanomyographic signals are shown in Figure 2Go. Balloon pressure measurements were very sensitive to artifacts; any movement close to the head caused significant interference. In addition, the baseline showed substantial movement because of small signal heights making analysis of signals with small amplitudes very difficult. In order to avoid artifacts, surgery needed to be stopped during the measurements. In two patients, the balloon needed careful re-attachment during the recovery period because the tape had disconnected from the skin.

Onset, maximal effect, T 25%, T 75%, and T 90% were similar for both methods (TableGo). Correlation between the balloon pressure method and phonomyography was very good (r = 0.98 for all T 1% twitches). Mean bias was 1% (LOA -9, +9 %) for all signals (Figure 3aGo). There was no significant difference in the measurement of the times to reach TOF-ratios of 0.5, 0.7, 0.8, and 0.9 (TableGo). Correlation between the balloon pressure method and phonomyography was very good with r = 0.99 for the times to reach TOF-ratios of 0.5, 0.7, 0.8 and 0.9. Mean bias for time to reach TOF-ratios of 0.5, 0.7, 0.8, and 0.9 was 0.1 min (LOA: -0.6 min, 0.8 min; Figure 3bGo). There was no drift (T 1% higher at full recovery than the initial value) monitored by the balloon pressure method and only a minimal drift with phonomyography (103 ± 3 %).


View this table:
[in this window]
[in a new window]
 
TABLE
 


View larger version (28K):
[in this window]
[in a new window]
 
FIGURE 3A Bland-Altman test for all evoked signals. Mean bias was 1% with limits of agreement of -9 and +9% for all signal.

 


View larger version (13K):
[in this window]
[in a new window]
 
FIGURE 3B Bland-Altman test for train of four ratios 0.5, 0.7, 0.8, 0.9. Mean bias was 0.1 min with limits of agreement of -6 and +8 min. MMG = balloon pressure mechanomyography; PMG = phonomyography.

 

    Discussion
 TOP
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
Onset and offset of neuromuscular blockade measured using phonomyography were very close to those obtained using the balloon pressure method. Agreement between balloon pressure mechanomyography and phonomyography was very good with minimal bias. Both methods can be used interchangeably to measure onset, offset and maximum effect of neuromuscular blockade after mivacurium at the corrugator supercilii muscle. Phonomyography showed minimal drift in this study, there was no drift with the balloon pressure mechanomyography.

The advantage of phonomyography lies in the fact that it is less difficult to apply, does not need special set-up procedures, and is less prone to artifacts. This method only needs a small condenser microphone (special in its ability to record very low frequency sounds), which is taped to the region above the eyebrow and is less prone to artifacts than the balloon pressure measurements. It is clear that balloon pressure mechanomyography is not a classical "mechanomyographic" measurement using a force transducer as it is available for the adductor pollicis muscle. However, it measures force, in analogy to cuff pressure measurements of the larynx which measures force of contraction of adducting laryngeal muscles. Prior to the study we experimented with balloons of different shapes and volumes in order to find a balloon which was most suitable to measuring force of contraction of the corrugator supercilii muscle. The balloon and the form of application that was used finally in this study produced the signals with highest amplitudes and lowest baseline movements: a balloon of 2.5 x 2.5 cm, rather square in shape with flattened sides that can be firmly taped to the skin. To restrict upward movement of the balloon away from the skin (and herewith underestimating the actual signal amplitude) a small additional piece of plastic, analogous to preload counterweights for classical mechanomyographic measurements of the adductor pollicis muscle, is attached firmly to bone structures of the face without movement. The balloon pressure method seemed to be too prone to artifacts and too much care needed to be taken in comparison to phonomyography for setup and careful monitoring; any movement disturbed the signal and caused slight movements. Subtle spontaneous contractions of the forehead that might occur in patients under lighter anesthesia disturb the balloon pressure monitoring. We would therefore favour phonomyography as a simple method which can be applied by just gluing a small microphone to the skin.

In a previous study,3 we have shown that phonomyography can easily be applied at the corrugator supercilii muscle, where its monitoring sensibility is better than those of acceleromyography. The fact that it can be applied at various muscles of interest in neuromuscular research (corrugator supercilii muscle, adductor pollicis muscle, adducting laryngeal muscles or diaphragm) and the good to very good agreement with methods measuring the actual force of muscle contraction (cuff pressure mechanomyography at the larynx,8 mechanomyography at the adductor pollicis muscle)9 makes it an interesting method for simultaneous monitoring of several muscles for which mechanomyography is difficult or impossible to apply.

There are still some problems with this method; since phonomyography measures low frequency sounds, artifacts are possible. Vessel pulsations can cause rhythmic sounds, which are detected as small waves of the baseline (Figure 4Go). However, these artifacts are easily recognizable and do not disturb the evoked measurements. Since the microphone is simply taped to the skin, care needs to be taken not to detach the tape and therefore lose the contact between the microphone and the skin surface. If this happens, the initial signal and the signal after recuperation of neuromuscular blockade are not necessarily of the same height or shape. Here, our group has begun technical improvements that may improve this situation by fitting the microphone with a self-adhesive surface, which then can be removed easily from the patient’s skin, similar to electrocardiogram electrodes.



View larger version (13K):
[in this window]
[in a new window]
 
FIGURE 4 Arterial pulsation as artifact.

 
So far, we have evaluated several aspects of phonomyography. A frequency domain analysis and determination of peak frequency to identify the technical properties of the method have shown that suitable microphones need to be able to record sounds between 2 Hz and 50 Hz.3 The duration of control stimulation does not change the measured onset, maximum effect or offset of neuromuscular blockade.11 The determination of the LOA between phonomyography and methods measuring the actual force of contraction has been very promising in showing narrow LOA. Our current research efforts deal with more technical questions as to whether the microphone size has an impact on sound detection and whether the size of the microphone influences measurement of neuromuscular blockade.

Balloon pressure mechanomyography correlates and agrees well with phonomyography under carefully-controlled conditions, but its baseline variability and susceptibility to artifacts from patient movements does render it impractical for routine clinical use. Phonomyography is an attractive method for routine clinical use because of its ease of application, baseline stability, and absence of elaborate devices.


    Footnotes
 
This work was performed using internal departmental funds.

Accepted for publication June 30, 2003. Revision accepted November 11, 2003.


    References
 TOP
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
1 Plaud B, Debaene B, Donati F. The corrugator supercilii, not the orbicularis oculi, reflects rocuronium neuromuscular blockade at the laryngeal adductor muscles. Anesthesiology 2001; 95: 96–101.[Medline]

2 Donati F, Plaud B, Meistelman C. A method to measure elicited contraction of laryngeal adductor muscles during anesthesia. Anesthesiology 1991; 74: 827–32.[Medline]

3 Hemmerling TM, Donati F, Beaulieu P, Babin D. Phonomyography of the corrugator supercilii muscle: signal characteristics, best recording site and comparison with acceleromyography. Br J Anaesth 2002; 88: 389–93.[Abstract/Free Full Text]

4 Barry DT. Muscle sounds from evoked twitches in the hand. Arch Phys Med Rehabil 1991; 72: 573–5.[Medline]

5 Barry DT, Cole NM. Muscle sounds are emitted at the resonant frequencies of skeletal muscle. IEEE Trans Biomed Eng 1990; 37: 525–31.[Medline]

6 Dascalu A, Geller E, Moalem Y, Manoah M, Enav S, Rudick Z. Acoustic monitoring of intraoperative neuromuscular block. Br J Anaesth 1999; 83: 405–9.[Abstract/Free Full Text]

7 Bellemare F, Couture J, Donati F, Plaud B. Temporal relation between acoustic and force responses at the adductor pollicis during nondepolarizing neuromuscular block. Anesthesiology 2000; 93: 646–52.[Medline]

8 Hemmerling TM, Babin D, Donati F. Phonomyography as a novel method to determine neuromuscular blockade at the laryngeal adductor muscles: comparison with the cuff pressure method. Anesthesiology 2003; 98: 359–63.[Medline]

9 Hemmerling TM, Babin D, Donati F. Phonomyography and mechanomyography can be used interchangeably to measure neuromuscular block at the adductor pollicis muscle. ASA Meeting Abstracts 2002; 987 (abstract).

10 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307–10.[Medline]

11 Hemmerling TM, Donati F, Babin D, Beaulieu P. Duration of control stimulation does not affect onset and offset of neuromuscular blockade at the corrugator supercilii muscle measured with phonomyography or acceleromyography. Can J Anesth 2002; 49: 913–7.[Abstract/Free Full Text]




This article has been cited by other articles:


Home page
Br J AnaesthHome page
A. A. Dahaba, H. Bornemann, B. Holst, G. Wilfinger, and H. Metzler
Comparison of a new neuromuscular transmission monitor compressomyograph with mechanomyograph
Br. J. Anaesth., March 1, 2008; 100(3): 344 - 350.
[Abstract] [Full Text] [PDF]


Home page
Canadian J. AnesthesiaHome page
T. M. Hemmerling and N. Le
Brief review: Neuromuscular monitoring: an update for the clinician: [Article de synthese court : Monitorage neuromusculaire : une mise a jour pour le clinicien]
Can J Anesth, January 1, 2007; 54(1): 58 - 72.
[Abstract] [Full Text] [PDF]


Home page
Canadian J. AnesthesiaHome page
G. Trager, G. Michaud, S. Deschamps, and T. M. Hemmerling
Comparison of phonomyography, kinemyography and mechanomyography for neuromuscular monitoring: [Comparaison de la phonomyographie, la cinemyographie et la mecanomyographie pour le monitorage neuromusculaire]
Can J Anesth, February 1, 2006; 53(2): 130 - 135.
[Abstract] [Full Text] [PDF]


Home page
Br J AnaesthHome page
S. Deschamps, G. Trager, P. A. Mathieu, and T. M. Hemmerling
The staircase phenomenon at the corrugator supercilii muscle in comparison with the hand muscles
Br. J. Anaesth., September 1, 2005; 95(3): 372 - 376.
[Abstract] [Full Text] [PDF]


Home page
Canadian J. AnesthesiaHome page
L. Olivieri and G. Plourde
Prolonged (more than ten hours) neuromuscular blockade after cardiac surgery: report of two cases: [Un blocage neuromusculaire prolonge (plus de dix heures) apres une intervention en cardiochirurgie : presentation de deux cas]
Can J Anesth, January 1, 2005; 52(1): 88 - 93.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Résumé de cet Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hemmerling, T. M.
Right arrow Articles by Donati, F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hemmerling, T. M.
Right arrow Articles by Donati, F.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS