| Abstract |
|---|
|
|
|---|
Methods: In a prospective observational study, we evaluated the BIS monitor in 87 children (ages: 0.3 to 14 yr) ASA physical status III undergoing general surgery under sevoflurane and nitrous oxide. Clinical signs of inhalational anesthesia (CSA), the motor response to surgical incision and signs of arousal were used as indicators of the depth of anesthesia. CSA and BIS measurements were paired every minute during induction and emergence until arousal.
Results: CSA scores decreased during induction and increased during emergence (P < 0.001) and correlated with changes in sevoflurane concentrations (r = 0.46; P < 0.001). BIS was associated with changes in CSA scores during induction (r = 0.49; P < 0.01) and emergence (r = 0.62; P < 0.01), but the ranges of individual BIS values overlapped several levels of hypnosis. A BIS value greater than 50 had a positive predictive value of 25% for distinguishing between responders and non-responders to surgical incision. A BIS score equal or greater than 72 had a predictive value of 63% for discriminating between pre-arousal and arousal.
Conclusions: BIS correlates with several levels of hypnosis during inhalational induction and emergence in children, but individual BIS values show large inter-individual variability. The BIS monitor identified the physiological changes associated with arousal and distinguished the effects of preoperative sedation during emergence. The use of movement as an endpoint of hypnosis during surgical stimulation does not correlate with BIS values in children. The large inter-individual variability of BIS at different levels of anesthetic depth may limit the applicability of BIS to pediatric anesthesia.
| Introduction |
|---|
|
|
|---|
The clinical validation of the BIS monitor during anesthesia in children requires correlating the BIS values with several levels of sedation or hypnosis, but this correlation may be associated with some limitations particularly in young children and infants.4,5 The behavioural response of young children to verbal commands is often not reliable. Also, infants are typically uncooperative.5,6 The evaluation of objective clinical signs7,8 that change uniformly during inhalational induction such as the breathing pattern, eye movements, diameter of pupils and their reactivity to light, may provide a more reliable and valid assessment of the transition from awake to sleep in children. In addition, in the absence of muscle relaxants, the movement response to a standard stimulus such as skin incision and the purposeful response or eye opening during arousal may represent another objective means for explaining the effects of anesthetics.810 In this study, we used these clinical signs as indicators of the changes in the level of hypnosis during sevoflurane anesthesia. We postulated that the BIS predicts an increasing depth of inhalational anesthesia during induction, and anticipates the manifestations of arousal during recovery from inhalational anesthesia.
| Material and methods |
|---|
|
|
|---|
Premedication, if necessary, was achieved by midazolam (0.5 mgkg1). All patients had inhalational induction and maintenance under sevoflurane and nitrous oxide in oxygen (50/50). There was no use of muscle relaxant. Anesthetic management was performed by one of the three anesthesiologists participating in the study, who were blinded to the BIS results and the clinical assessment.
Clinical assessment
After proper consensus among the participating investigators, a three point-score scale was developed based on previously described clinical signs of inhalational anesthesia (CSA).710 These clinical signs were classified into four categories (Table A, available as Additional Material at www.cja-jca.org) as follows: I) breathing pattern, II) eye movements, III) size of pupils and IV) pupillary response to light. In this study, it was assumed that patients with a maximum CSA score of 12 were awake, while those with a minimum score of 4 would be deeply anesthetized. Prior to the beginning of the study, two independent observers (R.A.R., S.D.) assessed the reliability of the scale in a group of children using evaluations every minute during induction and emergence until the childs spontaneous arousal. Arousal was defined when any of the following events first occurred: 1) purposeful movement of limbs, 2) facial grimace, 3) crying or phonating, or 4) spontaneous eye opening. Pre-arousal was defined as the last measurement prior to any manifestation of arousal. The end-tidal concentrations of sevoflurane and nitrous oxide, electrocardiogram, pulse oximetry, non-invasive blood pressure, capnography and temperature were monitored at every minute during induction and emergence.
Electrophysiologic assessment
In a subsequent phase of the study, we evaluated the association between the BIS and depth of anesthesia as indicated by the CSA scale. With the child awake, a disposable BIS-Sensor electrode (Aspect Medical Systems, Newton, MA, USA) was placed on the left forehead at locations corresponding to Fpz and F7 grounded to the zygomatic arch and the system connected to the BIS processor module (software version 1.03.11en; Space Labs Medical, Mississauga, ON, Canada). The smoothing window was set at 15 sec and the update rate at two seconds. The EEG signals were band-passed using filters between 0.25 Hz and 70 Hz. The impedance for each electrode was maintained below 5 K
. The average BIS value from the last two 15-sec periods at the end of every minute was included in the analysis. "BIS arousal" was defined as the average BIS value within the first minute after arousal. BIS with a quality index below 25% were rejected. Prior to induction, baseline values consisting of a two-minute EEG recording, blood pressure and heart rate were established. Subsequently, nitrous oxide in oxygen (50/50) was delivered through a pediatric mask and this was followed by stepwise increases of sevoflurane (e.g., 3%, 5%, 8%) at fixed time intervals (approximately 1520 sec) until loss of consciousness was achieved. At this time, a laryngeal mask airway (LMA) was inserted. Both electrophysiologic and clinical measurements were paired every minute from the start of induction until five minutes post-induction. For each measurement, clinical assessment was performed first, and the BIS value documented immediately after. Additional measurements recorded before and after the incision evaluated the response to surgical stimulation. The clinical response to incision was assessed by whether there was a gross, purposeful muscular movement of the childs head or extremities, twisting or jerking associated with the time of surgical incision. Some patients were administered a caudal epidural analgesic before incision with bupivacaine 0.25% (0.751.0 mLkg1) containing epinephrine (5 µgmL1). During surgery, the BIS values were concealed to the attending anesthesiologist. After completion of surgery and including application of dressings, the inhalational agents were discontinued, followed immediately by LMA removal and fresh oxygen (6 Lmin1) was delivered to the patient through a facial mask. The BIS and clinical assessments were assessed every minute from surgical completion to spontaneous arousal.
Data and statistical analyses
Variables were mostly non-normally distributed and results are presented as the median and quartile values (25th, 75th quartile). Blood pressure and heart rates during induction are expressed in percentages relative to pre-induction measurements, while those during emergence are correlated to the discontinuation of anesthesia. The use of relative rather than absolute values facilitated patients acting as their own control and minimized differences in age-related normative values and blood pressure cuff size. The changes on the BIS, heart rates, blood pressures and sevoflurane concentrations at different depths of anesthesia as indicated by the CSA scale were assessed by non-parametric Friedman tests. Wilcoxon-sign rank tests evaluated the differences between two related samples (e.g., pre-arousal and arousal), and the Spearman rank correlation determined the degree of association between BIS and clinical or hemodynamic measurements. Between-group differences (e.g., with and without preoperative sedation) were assessed by Kruskal-Wallis tests. The sensitivity and positive predictive value of BIS was calculated to determine its ability to distinguish between arousal and pre-arousal and between responders and non-responders to surgical stimulation. A P < 0.05 was considered significant. Our sample size was calculated under the assumption that a minimum of 50 patients would provide a 90% power (alpha = 0.05) to conclude that at least 80% of our population could show BIS values below 50 if the CSA scale decreased from 12 to 4.
| Results |
|---|
|
|
|---|
Clinical assessment
CSA scores decreased (P < 0.001) during induction and increased (P < 0.001) during emergence. The changes in clinical signs correlated with the end-tidal concentrations of sevoflurane (r = 0.46; P < 0.001). The CSA scale showed an acceptable inter-observer correlation during induction (r = 0.83; P < 0.001) and recovery (r = 0.93; P < 0.001). From all categories, the breathing pattern showed the best inter-observer correlation during induction (r = 0.90; P < 0.001), followed by eyes position (r = 0.84; P < 0.001), size of pupils (r = 0.78; P < 0.001) and pupillary reactivity to light (r = 0.77; P < 0.001). In contrast, during emergence the inter-observer correlation was better than induction for all categories (r = 0.85, 0.88, 0.87, 0.89; P < 0.001). Figure A (available as Additional Material at www.cja-jca.org) shows the variability in clinical assessments between the two observers.
Electrophysiologic assessment
EEG FEATURES
In the majority of children, the EEG during induction and recovery showed changes similar to those illustrated in Figure 1
. The EEG in an awake child (Figure 1
, panel A) was characterized by a dominant fast frequency (> 13 Hz) and low-amplitude (< 50 µV) activity. Induction typically started with an increase in beta-range (> 13 Hz) activity and immediately changed to a slow EEG activity (< 8 Hz) with dominant large amplitude (> 400 µV) slow delta waves (< 2 Hz); (Figure 1
, panel B). At deeper levels (Figure 1
, panel C), the EEG gradually changed to a mixed spectrum of medium frequency activity (610 Hz) with delta waves in the high-delta range (< 4 Hz). Typically, at CSA scores of 4, the EEG showed a steady-pattern of moderate amplitude (200250 µV) rhythmic delta activity (< 4 Hz) intermixed with diffuse theta (< 8 Hz > 4 Hz) waves (Figure 1
, panels D and E). Following discontinuation of anesthesia (Figure 1
, panels G, H), the delta wave activity disappeared and the EEG spectrum shifted from slow (< 8 Hz) to fast frequency activity (> 13 Hz). Near-arousal (Figure 1
, panel I), the EEG was characterized by a dominant fast activity with occasional bursts of diffuse theta (< 8 Hz > 4 Hz) waves. Finally, during arousal (Figure 1
, panel J), fast activity (> 13 Hz) and frontalis muscle activity were the main components of the EEG activity. Seven children (aged: 2 to 9 yr) showed transient episodes of EEG burst suppression at rates that varied between 30% and 72%. Two patients (ages: 2 and 5 yr) had a discharge of multiple spikes during induction followed by a period of transient burst suppression with rhythmic polyspikes. No abnormal movements were seen and their recovery was similar to the rest of the group.
|
|
|
|
| Discussion |
|---|
|
|
|---|
Several factors may be associated with the variability of BIS values in our study. Inhalational induction involved rapid changes of the inspired concentration of sevoflurane, which did not represent steady-state measurements. A previous study4 indicated that minimum alveolar concentrations (MAC) of sevoflurane are higher in infants compared to children. We observed frontalis muscle activity particularly at initial stages of induction, which has been described to potentially corrupt the BIS calculations.3 Some of our patients had periods of EEG burst suppression. A previous report14 suggests that the BIS algorithm during those episodes is insufficient to calculate accurately the BIS values and paradoxical increases in BIS values may be expected as a result of increases in anesthetic concentrations. We suspect that the time for computation of BIS scores is critical during induction in children. BIS scores resulting from longer computational times may not accurately reflect the faster changes in anesthetic concentration or clinical signs. Although there was no effect of age on BIS values, it is known that the EEG in the pediatric population shows maturational changes with age and that the effects of sedation on the EEG are particularly more pronounced in younger children.15 In addition, it is recognized that a smooth transition from awake to physiologic sleep may not occur in young children as compared with adults and this may result in a sudden drop into deep sleep with the consequent omission of transitional stages on the EEG.15,16
In our investigation, children who moved in response to surgical incision had pre-incision BIS values lower than 50 except for one (BIS: 63). It is expected that lower BIS values indicate a low probability of motor response during incision,3,17 but our findings suggest that low BIS values in children do not guarantee the absence of a motor response during surgical stimulation. Katoh et al.18 indicated that BIS was not a significant predictor of movement in response to skin incision under sevoflurane. It is assumed that the apparent failure of the BIS for predicting movement may be associated to differences in the anatomical sites responsible for the motor response (e.g., spine) as compared with the neural structures generating the EEG (e.g., cortex).2,3
There was a 9% incidence of movement in response to surgical stimulation. Although the rate appears to be high, we speculate that these findings may be related either to the non-caudal group not receiving supplemental analgesia or to the highly sensitive method of detection used in our study. The concomitant use of regional anesthesia in some of our patients introduced a confounding variable in the assessment of the motor response to surgical incision. None of our patients with caudal analgesia moved in response to skin incision. Moreover, no changes in BIS values were found between before and after the caudal administration. These results appear to be in agreement to those of Morley et al.19 and Ibrahim et al.11 who did not find any changes in the BIS values of adults following epidural anesthesia, but another study20 found a decreased anesthetic requirement following epidural blockade suggesting a moderate but significant effect on inhalational anesthetic requirements.
Two of our patients without history of neurologic symptoms had a discharge of multiple spikes during induction. Previous reports2123 indicate that this phenomenon may occur in a small number of adult and pediatric patients under sevoflurane induction. Although the factors involved with these events are not clear, it has been reported that adult patients who had controlled hyperventilation or faster induction are more likely to show these alterations compared with those under normal ventilation or slower induction.23
In our study, BIS increased from the discontinuation of inhalational agents to arousal. This was the consequence of disappearance of the slow waves and increase in the high frequency activity of the EEG (3040 Hz). Similarly, Degoute et al.24 found an increase in BIS during recovery of sevoflurane anesthesia in children, but their BIS values during arousal were slightly higher than those found in our study. The discrepancy may be related to the use of different scales for assessing the level of hypnosis and to differences in anesthetic requirements associated with the type of surgeries included in their study.
Our study has limitations. We used adult BIS sensor electrodes because a pediatric version was not available. We found that these electrodes are not suitable for children. It is possible that this may have accounted for misplacement of the electrodes in some of our cases and failure to display BIS values. There was variability attributed to the use of premedication.25 Our analysis stratified patients based on whether they received premedication, however, the effects of initial sedation in these patients cannot be distinguished from the sevoflurane effect. We did not evaluate BIS and CSA scores at steady-state anesthetic concentrations. Thus, any attempt to correlate these two variables with MAC is inappropriate. In addition, our patients were not followed postoperatively and this precluded correlating intraoperative BIS values with postoperative recall.
Our study incorporated previously described clinical signs of inhalational anesthesia because changes in the level of consciousness associated with inhalational induction are faster810 and these signs appear to be more consistent than the childs behavioural response. Our findings indicate that CSA scores had a fair correlation with end-tidal sevoflurane concentrations (r = 0.46). Although Cullen et al.7 used these clinical signs for halogenated agents different from sevoflurane, clinical eye signs at induction are described as reliable indicators of the level of anesthesia under sevoflurane.10 The latency between initial induction and changes in several eye signs appears to correlate with the different stages of sevoflurane induced sleep as determined by the EEG.10 Moreover, the maximum level of EEG depression has been associated with reductions in pupil size or reactivity, and apnea.26,27 Nevertheless, the applicability of our clinical assessment for agents different from sevoflurane will have to be determined.
In summary, BIS correlates with several levels of hypnosis during inhalational induction and emergence in children, but individual BIS values show large inter-individual variability. This variability may be influenced by several factors such as age, type of anesthesia, non-steady state anesthetic concentrations, preoperative sedation, time for computation of the BIS algorithm, frontalis muscle activity, degree of EEG burst suppression and clinical assessment. The BIS monitor identified the physiological changes associated with arousal and distinguished the effects of preoperative sedation during emergence. The use of movement as an endpoint of hypnosis during surgical stimulation does not correlate with BIS values in children. The large inter-individual variability of BIS at different levels of anesthetic depth may have implications for the applicability of BIS to pediatric anesthesia. Further studies correlating BIS and clinical endpoints of hypnosis are necessary prior to determine whether current BIS versions are applicable to pediatric anesthesia.
| Acknowledgments |
|---|
| Footnotes |
|---|
Accepted for publication September 16, 2003. Revision accepted January 14, 2004.
| References |
|---|
|
|
|---|
2 Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology 1998; 89: 9801002.[Medline]
3 Johansen JW, Sebel PS. Development and clinical application of electroencephalographic bispectrum monitoring. Anesthesiology 2000; 93: 133644.[Medline]
4 Denman WT, Swanson EL, Rosow D, Ezbicki K, Connors PD, Rosow CE. Pediatric evaluation of the bispectral index (BIS) monitor and correlation of BIS with end-tidal sevoflurane concentration in infants and children. Anesth Analg 2000; 90: 8727.
5 Davidson AJ, McCann ME, Devavaram P, et al. The differences in the bispectral index between infants and children during emergence from anesthesia after circumcision surgery. Anesth Analg 2001; 93: 32630.
6 Bannister CF, Brosius KK, Sigl JC, Meyer BJ, Sebel PS. The effect of bispectral index monitoring on anesthetic use and recovery in children anesthetized with sevoflurane in nitrous oxide. Anesth Analg 2001; 92: 87781.
7 Cullen DJ, Eger EI II, Stevens WC, et al. Clinical signs of anesthesia. Anesthesiology 1972; 36: 2136.[Medline]
8 Bimar J, Bellville JW. Arousal reactions during anesthesia in man. Anesthesiology 1977; 47: 44954.[Medline]
9 Epstein RH, Mendel HG, Guarnieri KM, Staudt SR, Lessin JB, Marr AT. Sevoflurane versus halothane for general anesthesia in pediatric patients: comparative study of vital signs, induction, and emergence. J Clin Anesth 1995; 7: 23744.[Medline]
10 Power C, Crowe C, Higgins P, Moriarty DC. Anaesthetic depth at induction. An evaluation using clinical eye signs and EEG polysomnography. Anaesthesia 1998; 53: 73643.[Medline]
11 Ibrahim AE, Taraday JK, Kharasch ED. Bispectral index monitoring during sedation with sevoflurane, midazolam, and propofol. Anesthesiology 2001; 85: 11519.
12 Berkenbosch JW, Fichter CR, Tobias JD. The correlation of the bispectral index monitor with clinical sedation scores during mechanical ventilation in the pediatric intensive care unit. Anesth Analg 2002; 94: 50611.
13 Nieuwenhuijs D, Coleman EL, Douglas NJ, Drummond GB, Dahan A. Bispectral index values and spectral edge frequency at different stages of physiologic sleep. Anesth Analg 2002; 94: 1259.
14 Detsch O, Schneider G, Kochs E, Hapfelmeier G, Werner C. Increasing isoflurane concentration may cause paradoxical increases in the EEG bispectral index in surgical patients. Br J Anaesth 2000; 84: 337.
15 Kellaway P. An orderly approach to visual analysis : characteristics of the normal EEG of adults and children. In: Daly DD, Pedley TA (Eds). Current Practice of Clinical Electroencephalography, 2nd ed. New York: Raven Press, Ltd.; 1990: 13999.
16 de Weerd AW, van den Bossche RA. The development of sleep during the first months of life. Sleep Med Rev 2003; 7: 17991.[Medline]
17 Sebel PS, Lang E, Rampil IJ, et al. A multicenter study of bispectral electroencephalogram analysis for monitoring anesthetic effect. Anesth Analg 1997; 84: 8919.[Abstract]
18 Katoh T, Suzuki A, Ikeda K. Electroencephalographic derivatives as a tool for predicting the depth of sedation and anesthesia induced by sevoflurane. Anesthesiology 1998; 88: 64250.[Medline]
19 Morley AP, Chung DC, Wong AS, Short TG. The sedative and electroencephalographic effects of regional anaesthesia. Anaesthesia 2000; 55: 8649.[Medline]
20 Pollock JE, Neal JM, Liu SS, Burkhead D, Polissar N. Sedation during spinal anesthesia. Anesthesiology 2000; 93: 72834.[Medline]
21 Komatsu H, Taie S, Endo S, et al. Electrical seizures during sevoflurane anesthesia in two pediatric patients with epilepsy. Anesthesiology 1994; 81: 15357.[Medline]
22 Yli-Hankala A, Vakkuri A, Sarkela M, Lindgren L, Korttila K, Jantti V. Epileptiform electroencephalogram during mask induction of anesthesia with sevoflurane. Anesthesiology 1999; 91: 1596603.[Medline]
23 Kaisti KK, Jaaskelainen SK, Rinne JO, Metsahonkala L, Scheinin H. Epileptiform discharges during 2 MAC sevoflurane anesthesia in two healthy volunteers. Anesthesiology 1999; 91: 19525.[Medline]
24 Degoute CS, Macabeo C, Dubreuil C, Duclaux R, Banssillon V. EEG bispectral index and hypnotic component of anaesthesia induced by sevoflurane: comparison between children and adults. Br J Anaesth 2001; 86: 20912.
25 Crain N, Slonim A, Pollack MM. Assessing sedation in the pediatric intensive care unit by using BIS and the COMFORT scale. Pediatr Crit Care Med 2002; 3: 114.[Medline]
26 Tayefeh F, Larson MD, Sessler DI, Eger EI II, Bowland T. Time-dependent changes in heart rate and pupil size during desflurane or sevoflurane anesthesia. Anesth Analg 1997; 85: 13626.[Abstract]
27 Nakatsuka I, Ochiai R, Takeda J. Changes in heart rate variability in sevoflurane and nitrous oxide anesthesia: effects of respiration and depth of anesthesia. J Clin Anesth 2002; 14: 196200.[Medline]
This article has been cited by other articles:
![]() |
K. S. Park, E. J. Hur, K. W. Han, H. Y. Kil, and T. H. Han Bispectral index does not correlate with observer assessment of alertness and sedation scores during 0.5% bupivacaine epidural anesthesia with nitrous oxide sedation. Anesth. Analg., August 1, 2006; 103(2): 385 - 9, table of contents. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Davidson, G. H. Huang, C. S. Rebmann, and C. Ellery Performance of entropy and Bispectral Index as measures of anaesthesia effect in children of different ages Br. J. Anaesth., November 1, 2005; 95(5): 674 - 679. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Shander and G. Lobel Caveats of Bispectral Index Monitoring in the Pediatric Population Chest, July 1, 2005; 128(1): 14 - 16. [Full Text] [PDF] |
||||
![]() |
I. Constant Utilisation du BIS en anesthesie pediatrique : ou en sommes-nous ?/BIS use in pediatric anesthesia: where are we? Can J Anesth, May 1, 2004; 51(5): 411 - 416. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |