| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |

* From the Department of Anesthesia and Critical Care, Landspitalinn University Hospital of Iceland, Reykjavik, Iceland; and
the Department of Anesthesiology, University of Wisconsin Medical School, Clinical Science Center, Madison, Wisconsin, USA.
Address correspondence to: Dr. Misha Perouansky, Department of Anesthesiology, University of Wisconsin Medical School B6/319 Clinical Science Center, 600 Highland Avenue, Madison, WI 53792-3272, USA. Phone: 608-263-4429; Fax: 608-262-5558; E-mail: mperouansky{at}wisc.edu
| Abstract |
|---|
|
|
|---|
Methods: We tested the Datex-Ohmeda 7900 ventilator mounted on an Excel 210 anesthesia machine under a variety of conditions. For comparison, we also tested the Ohmeda 7800 and the Dräger AV-2 ventilator under control conditions. All experiments were performed using a test lung.
Results: The oxygen consumption of the AV-2 and the Datex-Ohmeda ventilators averaged 302 ± 17 L·hr1 and 564 ± 68 to 599 ± 56 L·hr1, respectively (P < 0.01 AV-2 vs 7800 and 7900). When using an E-type cylinder, this would result in a mean time to alarm of 93 min and 54 to 57 min, respectively. Decreased lung compliance increased the oxygen consumption to 848 ± 16 L·hr1.
Conclusions: Machine-driven mechanical ventilation incurs a significant "oxygen cost." We show that the amount of oxygen consumed by mechanical ventilation with contemporary anesthesia ventilators is influenced by patient-dependent factors and may greatly exceed the amount of oxygen delivered to the patient.
| Introduction |
|---|
|
|
|---|
| Materials and methods |
|---|
|
|
|---|
The amount of O2 consumed to drive the bellows was determined as follows: the anesthesia machine was disconnected from the wall gas supply with all flow metres set to zero flow. A full E-type cylinder was installed. The FGF was set at 5 L O2·min 1 (to fill the circuit). Mechanical ventilation was initiated and, after two minutes, the FGF was reduced to 2 L O2·min1 for the rest of the experiment.
Tidal volumes (VT), peak, mean and minimal circuit pressures were continuously monitored with the builtin spirometer and confirmed with the test lung at least twice during the experiment. If set and measured VTs differed by more than 10% the experiment was discontinued.
Exactly 30 min after the start of the experiment the anesthesia machine was turned off, the O2 cylinder closed and released from the anesthesia machine. The total volume of O2 (VTOT) consumed during the experiment was determined from the difference in the mass of the cylinder before and after the experiment using the ideal gas-equation VTOT = [
m x [(273 + (Tamb)/273] x 22.4 L x mol 1)]/32 g x mol1,1 where
m = weight of cylinder in grams at the beginning minus weight after the experiment, (Tamb) = numerical value of measured ambient temperature in degrees cel-cius, 22.4 L and 32 g are the volume at 0°C and the weight of one mole of O2. The scales used to measure the cylinders weight were precise to ± 1 g (± 0.75 L O2 at 20°C). The O2 requirements of the ventilator (Vvent) over the 30 min of experiment were calculated by subtracting the FGF from VTOT: Vvent = VTOT - (5 L·min1 x 2 min + 2 L·min1 x 28 min) and converted to litres per hour. Oxygen requirements were compared using the non-paired t test, a P-value of < 0.05 was considered significant.
For experiments under standard conditions, the ventilator was set to deliver a VT of 700 mL at a rate of 10 breaths·min1. The delivered volume was tested with the test lung and the ventilator settings adjusted to deliver a measured VT of 700 ± 70 mL. This was done with the anesthesia machine connected to central gas supply. Test lung compliance and airway resistance were set at 50 mL/cm H2O and 20 cm H2O/L·min1 (normal values from BIO-TEK Instruments Inc., Winooski, VT, USA), respectively. For some experiments with the 7900 these variables were adjusted on the test lung as follows: in order to simulate lungs with high and low compliances (e.g., open chest vs adult respiratory distress syndrome) the compliance of the artificial lung was set at 100 and 20 mL/cm H2O, respectively. For experiments involving pressure-controlled ventilation the inspiratory pressure was adjusted to deliver a VT of 700mL as measured by the test lung. We performed a total of 41 experiments. The number of experiments performed with each ventilator ranged from three to five for each condition tested.
| Results |
|---|
|
|
|---|
Control conditions
The measured O2 requirements of the tested ventilators under control conditions are presented in Table I
. The Ohmeda ventilators required significantly more O2 to drive the ventilator bellows than the Narkomed. Under control conditions, the 7900 and 7800 ventilators required approximately twice as much O2 as the AV-2 (599 ± 56 and 564 ± 68 vs 302 ± 17 L·hr1, P < 0.01). Selecting pressure-controlled ventilation instead of volume-controlled ventilation to deliver the same minute volume did not change the driving gas requirements of the 7900 ventilator (598 ± 30 L·hr1).
|
|
|
| Discussion |
|---|
|
|
|---|
Under control conditions (VT 700 mL, 10 breaths·min1, compliance 50 mL/cm H2O, PEEP 0), the Ohmeda ventilators 7800 and 7900 had similar O2 requirements ranging from 564 to 599 L·hr1 for the 7800 and the 7900, respectively (P = 0.45). The Dräger AV-2 (with the inspiratory flow rate set to medium) required approximately 50% less O2 than the Ohmeda ventilators (i.e., 302 ± 17 L·hr1). A difference in O2 consumption was expected: during mechanical ventilation the AV-2 (like other North American Dräger ventilators) entrains air, exploiting the Venturi effect, and therefore drives the bellows with a gas mixture containing ± 33% entrained air when inspiratory flows are set to mid-range (North American Dräger, personal communication),3 while Ohmeda ventilators compress the bellows with 100% O2. Our experiments show that using a rule of thumb (e.g., driving gas = minute ventilation) generally underestimates the actual driving gas requirements of the ventilators. Under baseline conditions we ventilated with a minute ventilation of 7 L, i.e., 420 L·hr1 (excluding FGF). This would result in an estimated driving gas requirement of ± 420 L and ± 280 L of O2 per hour for the Ohmeda and the Dräger ventilators (assuming 33% entrained air in the driving gas for the latter): an underestimate of the actual consumption by 40% and 8%, respectively. Our data indicate that the error would be more significant when ventilating diseased lungs with low compliance.
Modern anesthesia ventilators offer the practitioner a choice between volume- and pressure-controlled ventilation. We examined whether driving gas requirements of the 7900 were affected by the choice of a specific ventilation mode. Under control conditions, there was no significant difference in driving gas requirements between volume-controlled and pressure-controlled ventilation (599 ± 56 vs 598 ± 30 L·hr1).
Finally, we defined the impact of varying lung compliance on driving gas requirements of the 7900. Decreasing lung compliance increased the driving gas requirements (Figure
), but the difference reached statistical significance only for the most extreme case tested - a compliance of 20 mL/cm H2O, still within the range encountered clinically. Therefore, the patients lung compliance should be taken into account, even though its impact on O2 consumption would probably remain unnoticed under usual clinical conditions. We noticed that the variability of the O2 requirements of the 7900 increased with decreasing compliance of the test lung. As the experiments were performed on the same machine, variability between ventilators cannot account for this observation. While we do not have a simple explanation, we think it is possible that the microprocessor-controlled feedback circuitry of the 7900, that provides constant adjustments of delivered VTs, might oscillate with increasing impedance of the breathing circuit. It is remarkable that application of PEEP reduced the variability dramatically (Figure
).
Our results for ventilation under normal conditions are consistent with those obtained by Taenzer et al.2 In addition, our data can also be used to estimate the length of time that ventilation can be maintained with any size cylinder, as long as the amount of stored O2 is known.
In conclusion, we have measured the driving gas requirements of three contemporary anesthesia machine-mounted ventilators. In order to deliver 1 L of minute ventilation to the patient under control conditions, the Datex-Ohmeda 7800/7900 and Drägers AV-2 ventilators required approximately 1.4 L and 0.69 L of O2 driving gas, respectively. Ventilation of lungs with a low compliance increased the O2 requirements significantly: in order to deliver 420 L of minute ventilation, the 7900 required ± 848 L of O2, that is more than 2 L of driving gas per litre of minute ventilation. This translates into a reduction of the time an E-type cylinder would support ventilation from 54 to 39 min.
| Acknowledgments |
|---|
| Footnotes |
|---|
Accepted for publication December 8, 2003. Revision accepted March 26, 2004.
| References |
|---|
|
|
|---|
2 Taenzer AH, Kovatsis PG, Raessler KL. E-cylinder-powered mechanical ventilation may adversely impact anesthetic management and efficiency. Anesth Analg 2002; 95: 14850.
3 Cicman J, Himmelwright C, Skibo V, Yoder J. Operating Principles of Narkomed Anesthesia Systems. Telford, PA: North American Dräger; 1993: 189.
4 Dorsch JA, Dorsch SE. Understanding Anesthesia Equipment, 3rd ed. Philadelphia, PA: Lippincott Williams & Wilkins; 1998.
This article has been cited by other articles:
![]() |
D. F. Szpisjak, C. L. Lamb, and K. D. Klions Oxygen Consumption with Mechanical Ventilation in a Field Anesthesia Machine Anesth. Analg., June 1, 2005; 100(6): 1713 - 1717. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A.E. Shephard The changing pattern of anesthesia, 1954-2004: a review based on the content of the Canadian Journal of Anesthesia in its first half-century: [La transformation du modele de l'anesthesie, 1954-2004 : une revue fondee sur le contenu du premier demi-siecle du Journal canadien d'anesthesie] Can J Anesth, March 1, 2005; 52(3): 238 - 248. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |