sophagienne de la mise en place dun cathéter veineux central guidée par ECG]


* From the Departments of Anesthesiology, Seoul National University Hospital, Seoul; the
Seoul National University Bundang Hospital, Seongnam; and the
Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Korea.
Address correspondence to: Dr. Jae-Hyon Bahk, Department of Anesthesiology, Seoul National University Hospital, Seoul National University College of Medicine, 28 Yongon-dong, Jongno-gu, Seoul 110-744, Korea. Phone: 82-2-2072-2818; Fax: 82-2-747-5639; E-mail: bahkjh{at}snu.ac.kr
| Abstract |
|---|
|
|
|---|
Methods: We evaluated 54 patients for whom central venous catheterization was planned as part of routine care for their elective surgery. The junction of the superior vena cava (SVC) and the right atrium (RA) was defined as the superior border of the crista terminalis by transesophageal echocardiography. The RA ECGs were recorded while withdrawing the CVC into the SVC or advancing it into the RA at 1-cm intervals. Saline was used as an electrical conductor via the distal lumen of the CVC.
Results: The tallest peaked and biphasic P waves [median (interquartile range)] were observed when the CVC tip was located at positions 0.0 cm (1.0 to 0.0) and 4.0 cm (5.0 to 3.0) below the SVC/RA junction, respectively. The P wave returned to a normal shape and size at 4.0 cm (3.0 to 4.0) above the SVC/RA junction. Overshoot P waves were observed at 4.0 cm (5.0 to 3.0) below the SVC/RA junction in 22 patients, when the CVC tip appeared to be contacting or in close proximity to the RA wall.
Conclusions: During ECG-guided central venous catheterization, the tallest peaked P wave may be used to place the CVC tip at the SVC/RA junction, the normally-shaped P wave identifies the mid to upper SVC, and biphasic P waves identify RA localization.
| Introduction |
|---|
|
|
|---|
Unfortunately, clinicians have lacked reliable surface landmarks to position the CVC accurately without diagnostic testing. For anesthesiologists, routine use of fluoroscopy or post-procedure portable x-rays would lengthen operative time and increase costs. Additionally, such routine testing would expose operating room personnel and patients to extra irradiation. Using electrocardiography (ECG)-guided central venous catheterization, radiation hazards can be minimized and the costs of CVC placement can be reduced.5
Electrocardiography-guided central venous catheterization has been reported to be accurate and successful. 68 However, variations in the methodology exist, and the results evaluated radiologically appear to be inconsistent.5,711 Chu et al.12 have evaluated ECG-guided central venous catheterization using transesophageal echocardiography (TEE). However, this study was performed while analyzing only the tallest peaked P wave. There are no studies which describe the exact position of CVCs when the associated P wave displays are either biphasic or normal in shape. We therefore undertook a prospective descriptive clinical study using TEE guidance to establish the absolute locations and ranges of CVC tip positions when specific P wave patterns are displayed, and, thereby, to provide more accurate assessment of CVC localization during ECG-guided central venous catheterization.
| Methods |
|---|
|
|
|---|
In the operating room routine monitors were applied. Following induction of general anesthesia, the patients trachea was intubated and the lungs were mechanically ventilated to maintain normocarbia. A multi-plane TEE probe (Sonos 5500 imaging system, Philips, Andover, MA, USA) was placed into the patients esophagus and a bicaval view was obtained. All cannulations were performed by two anesthesiologists (Y.-S. Jeon and J.-H. Kim) with more than eight years experience in landmark-guided internal jugular vein access. Because only the right-side internal jugular vein was cannulated, the patient was placed in Trendelenburg position with the head turned to the left and the neck extended. Using standard technique, the needle was directed toward the ipsilateral nipple. A 30-cm-long, double- or triple-lumen CVC (B Braun, Mesungen, Germany) was advanced using a sterile Seldinger technique. After cannulation, the patient was returned to a horizontal plane.
During ECG-guided central venous catheterization, normal saline in the distal lumen of the CVC was used as the intravascular electrode to conduct electrical signals from the heart. The leads were attached to a standard three-lead ECG system. Lead II was monitored with the right arm lead (the negative electrode), the left leg lead (the positive electrode) and the left arm lead (the ground). The right arm electrode in standard lead II configuration was attached to a hypodermic needle passed through the injection port of a male Luerlock injection cap on the distal lumen of the CVC. The distal lumen of the CVC was carefully aspirated to remove any air bubbles and then flushed with sterile heparinized saline. The inserted hypodermic needle was connected to the right arm electrode of the ECG monitor using a pre-sterilized electric wire with alligator clamps at both ends, creating a catheter-to-left-leg lead. Real-time monitoring of the ECG tracing was conducted at the bedside with a portable 12 lead monitor terminal (MAC 5000; Marquette, Milwaukee, WI, USA).
Next, under TEE guidance on a bicaval view, the CVC was advanced until the tip of CVC reached the SVC/RA junction. Echocardiographically, this was defined as the base of the superior edge of the crista terminalis.13 To localize the CVC tip at the junction accurately, the bicaval and short axis views, longitudinal and right-angled to the CVC axis, were alternated to observe the distal CVC tip on the TEE frame while moving the TEE probe backwards and forwards. Agitated saline was flushed through the distal end of the CVC to facilitate identification of the tip.
When the CVC tip was positioned at the SVC/RA junction, the ECG was recorded on a portable 12-lead monitor. The CVC was advanced at 1-cm intervals until it reached 5 cm below the SVC/RA junction. The CVC was then withdrawn to the SVC/RA junction and the P wave appearance was rechecked. Thereafter, the CVC was withdrawn at 1-cm intervals to 7 cm above the SVC/RA junction. Each ECG tracing (lead II) was recorded at 1-cm intervals from 5 cm below the SVC/RA junction to 7 cm above it.
The tallest peaked P wave was defined at the peak amplitude of the P wave. The P wave was considered biphasic whenever the amplitude of the negative deflection approximated half the amplitude of the later positive P wave deflection, and the negative deflection was more than 1 mm in width at a standard ECG sweep speed of 25 mm·sec1 (Figure 1
). With the CVC tip below the SVC/RA junction, the overshoot P wave was defined when the amplitude of the P wave abruptly increased in both directions exceeding the amplitude of QRS wave (Figure 1
). While withdrawing the CVC, we verified whether the P wave appeared to have a normal size and shape compared with the P wave of conventional lead II. A normally-shaped P wave was defined by consensus between at least two of the three attending anesthesiologists. On the ECG tracing print, the vertical distances from the SVC/RA junction to the CVC tip were calculated to the nearest 1-cm interval, and were recorded at each display of the normally-shaped P wave, the tallest peaked P wave, the characteristic biphasic P wave and the overshoot P wave. Zero position referred to localization at the SVC/RA junction. Positive values refer to positions in the SVC above the SVC/RA junction, and negative values to positions in the RA below it.
|
| Results |
|---|
|
|
|---|
The tallest peaked P waves (n = 54) {median (interquartile range) [range]} were observed when the CVC tips were located at 0.0 cm (1.0 to 0.0) [3.0 to 0.0] below the SVC/RA junction. The P wave {median (interquartile range) [range]} returned to the normal upright configuration 4.0 cm (3.0 to 4.0) [2.0 to 5.0] above the SVC/RA junction (Figure 2
).
|
|
| Discussion |
|---|
|
|
|---|
Analysis of P wave morphology as a marker ECG-guided central venous catheterization has been evaluated previously. However, most of these studies report variable results, and do not include clear anatomical correlates. In a study by McGee et al.,10 the point of maximal P wave amplitude was presumed to be the catheter tip position closest to the sinoatrial (SA) node and the catheter was subsequently withdrawn 3 cm to locate the tip just proximal to the SVC/RA junction. In a study by Madan et al.,7 after the biphasic P wave had been observed, the catheter was withdrawn to obtain a peaked P wave to position the CVC tip in the lower SVC. In another study by Wilson et al.,6 after obtaining a biphasic P wave (defined as that measured when the catheter tip was opposite the SA node), the catheter was withdrawn until the P wave configuration changed to a peaked wave. In the study by Salmela et al.,11 after an augmented peaked P wave was observed (indicating that the tip of the CVC lay in the RA), the catheter was slowly withdrawn until the P wave began to diminish, and then withdrawn another 12 cm to lie in the SVC. Other investigators have recommended that to locate the CVC tip within the SVC,1417 the CVC catheter should be withdrawn gradually to the point at which the P wave is normalized, after a peaked P wave has been obtained.
Inconsistency in ECG-guided central venous catheterization and interpretations may be caused by the technical difficulty in evaluating the exact location of the CVC tip. In previous studies, the location of the CVC was confirmed by chest x-ray (CXR).5,6,11,14 Because a radiological definition of ideal location is not established, postoperative radiological evaluation may not have been performed accurately. On CXR, an apparent intersection of the SVC with the right superior heart border is often created by the left atrium instead of the RA.18 The relative location of the CVC compared with the RA/SVC junction tends to appear deeper than the actual location when using a portable anterior-posterior CXR.18 Furthermore, the position of the CVC tip may change during radiographic preparation because it may be displaced up to 1.5 to 3.0 cm with flexion and extension of the neck.19 In addition, radiological interpretations are often imprecise and subject to inter-observer variability, because different radiographic landmarks are used.6,10,20 However, the results for the tallest peaked P wave are consistent with previous studies using TEE12,17,21 or direct confirmation at the operating field.14 In this study, we believe that TEE facilitated assessment of CVC tip localization using a three-dimensional assessment of their positions.
Unsuccessful experiences have been reported when the CVCs are either too short, or form a loop.11 False negatives have been observed in patients with cardiac pacemakers, and those with cardiomyopathic dilatation or atrial fibrillation.5,11 If the CVC tip enters the internal jugular vein, the opposite subclavian vein, or the internal mammary vein,5,7 the ECG pattern is attenuated, and the P wave will not show any marked negative deflection with a subsequent biphasic pattern as CVCs are advanced: this occurred in 18 of 1,236 cases studied.5 In our study, when an overshoot P wave appeared, the characteristic biphasic P wave could be obtained by rotating the CVC in five of 22 patients. Although not characteristic, the overshoot P wave also has negative and later positive deflections, and can therefore be regarded as a type of biphasic P wave. However, such biphasic P wave patterns, including the characteristic and overshoot P waves, were not obtained in six of 54 patients. We do not know the exact causes, but one factor may have been a study population of primarily cardiac surgery patients, frequently with large RAs. Although we excluded patients with abnormal P waves, one limitation of this study is that 46 of the 54 patients enrolled were cardiac surgery patients, whose hearts were not normal.
During ECG-guided central venous catheterization, the biphasic P wave may be used to place the CVC tip in the RA. However, the characteristic biphasic P wave may not be obtained when the CVC tip is contacting or in close proximity to the RA wall. Although the overshoot P wave can be used to guide the CVC into the RA because the overshoot P wave is also biphasic, the CVC tip should not be located where the overshoot P wave is shown. The possibility of RA perforation may exist because of the fact that an overshoot P wave was observed only when CVC tips were making contact or in close proximity to the RA wall on TEE.
In a previous report,22 the mean SVC length in vivo was approximately 6 cm with almost half the SVC within the pericardium. Therefore, a normally-shaped P wave, which can be obtained from about 4.0 cm above the SVC/RA junction, can be used to guide the CVC tip at the mid to upper SVC, preventing the possibility of cardiac tamponade caused by CVC. However, if avoiding upper SVC positioning for a potential risk of thrombosis, the operator may use a small peaked P wave as a guide to locate the CVC tip in the lower SVC.
In conclusion, during ECG-guided central venous catheterization, the tallest peaked P wave may be used to place the CVC tip at the SVC/RA junction, the normally-shaped P wave identifies the mid to upper SVC, and a biphasic pattern of the P wave can be used to locate the RA.
| Footnotes |
|---|
Accepted for publication May 15, 2006. Revision accepted July 4, 2006.
| References |
|---|
|
|
|---|
2 Hanna PG, Gravenstein N, Pashayan AG. In vitro comparison of central venous catheters for aspiration of venous air embolism: effect of catheter type, catheter tip position, and cardiac inclination. J Clin Anesth 1991; 3: 2904.[Medline]
3 Schutz JC, Patel AA, Clark TW, et al. Relationship between chest port catheter tip position and port malfunction after interventional radiologic placement. J Vasc Interv Radiol 2004; 15: 5817.[Medline]
4 National Kidney Foundation. K/DOQI Clinical practice guidelines for vascular access. Am J Kidney Dis 2001; 37(Suppl 1): S13940.
5 Watters VA, Grant JP. Use of electrocardiogram to position right atrial catheters during surgery. Ann Surg 1997; 225: 16571.[Medline]
6 Wilson RG, Gaer JA. Right atrial electrocardiography in placement of central venous catheters. Lancet 1988; 27: 4623.
7 Madan M, Shah MV, Alexander DJ, Taylor C, McMahon MJ. Right atrial electrocardiography: a technique for the placement of central venous catheters for chemotherapy or intravenous nutrition. Br J Surg 1994; 81: 16045.[Medline]
8 Parigi GB, Verga G. Accurate placement of central venous catheters in pediatric patients using endocavitary electrocardiography: reassessment of a personal technique J Pediatr Surg 1997; 32: 12268.[Medline]
9 Martin JT. Neuroanesthetic adjuncts for patients in the sitting position. III. Intravascular electrocardiography. Anesth Analg 1970; 49: 793808.
10 McGee WT, Ackerman BL, Rouben LR, Prasad VM, Bandi V, Mallory DL. Accurate placement of central venous catheters: a prospective, randomized, multicenter trial. Crit Care Med 1993; 21: 111823.[Medline]
11 Salmela L, Aromaa U. Verification of the position of a central venous catheter by intra-atrial ECG. When does this method fail? Acta Anaesthesiol Scand 1993; 37: 268.[Medline]
12 Chu KS, Hsu JH, Wang SS, et al. Accurate central venous port-A catheter placement: intravenous electrocardiography and surface landmark techniques compared by using transesophageal echocardiography. Anesth Analg 2004; 98: 9104.
13 Andropoulos DB, Stayer SA, Bent ST, et al. A controlled study of transesophageal echocardiography to guide central venous catheter placement in congenital heart surgery patients. Anesth Analg 1999; 89: 6570.
14 Schummer W, Schummer C, Schelenz C, et al. Central venous cathetersthe inability of intra-atrial ECG to prove adequate positioning. Br J Anaesth 2004; 93: 1938.
15 Hoffman MA, Langer JC, Pearl RH, et al. Central venous catheters--no X-rays needed: a prospective study in 50 consecutive infants and children. J Pediatr Surg 1988; 23: 12013.[Medline]
16 Luks FI, Picard DL, Pizzi WF. Electrocardiographic guidance for percutaneous placement of central venous catheters. Surg Gynecol Obstet 1989; 169: 1578.[Medline]
17 Schummer W, Herrmann S, Schummer C, et al. Intraatrial ECG is not a reliable method for positioning left internal jugular vein catheters. Br J Anaesth 2003; 91: 4816.
18 Aslamy Z, Dewald CL, Heffner JE. MRI of central venous anatomy. Implications for central venous catheter insertion. Chest 1998; 114: 8206.
19 Curelaru I, Linder LE, Gustavsson B. Displacement of catheters inserted through internal jugular veins with neck flexion and extension. A preliminary study. Intensive Care Med 1980; 6: 17983.[Medline]
20 Artru AA, Colley PS. Placement of multiorificed CVP catheters via antercubital veins using intravascular electrocardiography. Anesthesiology 1988; 69: 1325.[Medline]
21 Schummer W, Schummer C, Schelenz C, Schmidt P, Fröber R, Hüttemann E. Modified ECG-guidance for optimal central venous catheter tip positioning. A transesophageal echocardiography controlled study (German). Anaesthesist 2005; 54: 98390.[Medline]
22 Kwon TD, Kim KH, Ryu HG, Jung CW, Goo JM, Bahk JH. Intra- and extra-pericardial lengths of the superior vena cava in vivo: implication for the positioning of central venous catheters. Anaesth Intensive Care 2005; 33: 3847.[Medline]