| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
From the Department of Anesthesiology, Faculty of Medicine, The University of Tokyo, Tokyo, Japan.
Address correspondence to: Dr. Tomoki Nishiyama, 3-2-6-603, Kawaguchi, Kawaguchi-shi, Saitama, 332-0015, Japan. Phone: 81-3-5800-8668; Fax: 81-3-5800-9655; E-mail: nishit-tky{at}umin.ac.jp
| Abstract |
|---|
|
|
|---|
Methods: Ten male volunteers were enrolled in this study after Institutional approval and obtaining informed consent. The probe of the Dolphin 2100, Nellcor N-595, or Masimo SET radical version 4.2 was attached to the left index finger. Time from power on to acquire the pulse wave and oxygen saturation (SpO2), time from the application of air tourniquet with 250 mmHg on the upper arm to loss of pulse wave and SpO2, and time from the release of the tourniquet to acquire the pulse wave and SpO2 were measured. Then, the patients left hand and arm were cooled gradually to 27°C dermal temperature in a room at 19°C. The temperatures at loss of the pulse wave and SpO2 were recorded.
Results: The Nellcor N-595 was the slowest to detect SpO2 and pulse wave at power on. The Masimo SET showed pulse wave and SpO2 longer than the other two monitors after tourniquet on. The Nellcor N-595 was the fastest to show pulse wave and SpO2 following tourniquet release.
Conclusion: The Masimo SET was the slowest to respond to the changes in perfusion, and the Nellcor N-595 responded the fastest. However, the Nellcor N-595 was the slowest to show SpO2 and pulse wave at power on.
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Data are shown as mean ± standard deviation. Statistical analysis was performed with the factorial analysis of variance followed by Student-Newman-Keuls test as a post-hoc test. A P value < 0.05 was considered to be statistically significant. A post-hoc power analysis was performed using the G power version 2.1.2 (University of Trier, Trier, Germany) with
= 0.05, f = 0.25 for each measurement (total sample size = 30 for three groups).
| Results |
|---|
|
|
|---|
|
| Discussion |
|---|
|
|
|---|
The Masimo SET might be more resistant than other monitors to the altered perfusion induced by a tourniquet. However, the Nellcor N-595 responded faster than the others when perfusion returned, and the Masimo SET was the slowest to recover. Application of tourniquet pressure at 250 mmHg might induce complete occlusion of blood flow rather than hypoperfusion in patients with normal blood pressure. Therefore, it might be possible that the response to loss of blood flow and to reflow was the slowest in the Masimo SET, which enabled the Masimo SET to show the pulse wave and SpO2 longer than the other monitors, even when blood flow was occluded. This does not imply that the Masimo SET could detect the pulse wave and SpO2 better than the other two systems at a reduced state of perfusion. Different blood pressures might influence the effects of the tourniquet. Although blood pressure was not measured in the present study, volunteers were young without hypertension, and each device was repeated for each subject. Therefore, the effects of blood pressure should be small.
Cooling the hand to 27°C, which was the lower limit in awake volunteers, had no effect on detecting pulse wave and SpO2 in each monitor tested. However, it was unclear how the perfusion decreased at 27°C, though fingers were cyanosed. Methods using a tourniquet or cooling might be different from clinical situations with hypoperfusion, such as hypovolemia, deep anesthesia or shock secondary to sepsis etc., because humoral or neural responses might also bear an influence in such clinical situations.
There are several studies comparing the Masimo SET with conventional pulse oximeters. Using the cooling environment and tapping and rubbing motions, the Masimo SET performed significantly better during motion and hypoperfusion than the other pulse oximeters (Nellcor N-395, Tyco Healthcare, Pleasanton, CA, USA; Datex-Ohmeda AS/3, Datex-Ohmeda, Madison, WI, USA; Marquette 8000, GE Healthcare IT, Milwaukee, WI, USA).4 In a study comparing the Masimo SET, Nellcor N-395, N-20PA, and D-25, time to loss of signal during blood pressure measurement by cuff was longer in the Masimo SET.3 The Masimo SET and the Nellcor N-395 showed shorter times to recover than the other monitor types.3 These studies suggest that the Masimo SET could offer advantages over conventional pulse oximeters. However, there have been no studies comparing the Masimo SET with other pulse oximeters using the new technologies.
Conventional pulse oximetry derives SpO2 from the ratios of emitted vs transmitted light (Lambert-Beers law), whereas the Masimo SET utilizes a patented discrete saturation transform algorithm to process the SpO2.2 The Masimo SET calculates SpO2 without first referencing the pulse rate, whereas recognition of a stable pulse is prerequisite to conventional oximeters. Therefore, response time was faster with the Masimo SET than the conventional oximeters. The Masimo SET pulse rate and SpO2 algorithms are so robust that they capture more true bradycardiac and hypoxemic events than conventional pulse oximeters during both motion and hypoperfusion.5 The Nellcor N-595 has a digital tip in the sensor (OxiMaxTM, Tyco Healthcare, Pleasanton, CA, USA).6 Each sensor of the OxiMaxTM contains all the calibration and operating characteristics for that individual sensor. It allows an information exchange between the sensor and the monitor, improving monitoring performance.6 Therefore, the Nellcor N-595 had a faster response than the Masimo SET. The Dolphin 2100 uses oximetry noise elimination (ONETM, Dolphin Medical, Hawthorne, CA, USA) technology, in which light signals are digitalized in the sensor. The different responses at hypoperfusion or occlusion of blood flow of these three monitors could not be related to the differences in technologies, due to lack of basic studies and details from the manufacturers. To better discriminate the clinical differences of these pulse oximeters, further studies using quantitative motion, hypoperfusion, and deeper cooling should be performed.
In conclusion, for the monitors tested, the Masimo SET was the slowest to respond to changes of perfusion, and the Nellcor N-595 responded the fastest. The Nellcor N-595 was the slowest to show SpO2 and pulse wave at power on.
| Footnotes |
|---|
Assessed April 11, 2005. Revision accepted for publication August 23, 2005. Final revision accepted September 8, 2005.
| References |
|---|
|
|
|---|
2 Dumas C, Wahr JA, Tremper KK. Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room. Anesth Analg 1996; 83: 26972.[Abstract]
3 Kawagishi T, Kanaya N, Nakayama M, Kurosawa S, Namiki A. A comparison of the failure times of pulse oximeters during blood pressure cuff-induced hypoperfusion in volunteers. Anesth Analg 2004; 99: 7936.
4 Barker SJ. Standardization of the testing of pulse oximeter performance. Anesth Analg 2002; 94(1 Suppl): S1720.[Medline]
5 Goldstein MR. Left heart hypoplasia. A life saved with the use of a new pulse oximeter technology. Neonatal Intensive Care 1998; 12: 147.
6 Mannheimer PD, Bebout DE. The OxiMax System. Nellcors new platform for pulse oximetry. Minerva Anestesiol 2002; 68: 2369.[Medline]
This article has been cited by other articles:
![]() |
P. B. Batchelder and D. M. Raley Maximizing the Laboratory Setting for Testing Devices and Understanding Statistical Output in Pulse Oximetry Anesth. Analg., December 1, 2007; 105(6S_Suppl): S85 - S94. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |