| Abstract |
|---|
|
|
|---|
Methods: One hundred and fifty patients were randomized into three groups. An identical general anesthesia technique was used in all patients apart from the addition of muscle relaxants and reversal drugs in the TT group. Patients were excluded if there were risk factors for gastroesophageal reflux. Ease of use, quality of seal, ventilation, systolic pressure, response to intubation, side effects and operating room time were assessed.
Results: Both ProSeal LMATM and SLIPATM were easy to insert (100% success) and ventilate with respective mean (standard deviation) maximum sealing pressures of 31 (4.6) and 30 (5.2) cmH2O (P = 0.4) with no muscle relaxants. The seal quality in both ProSeal LMATM and SLIPATM permitted the use of low flows, 485 (291) and 539 (344) mL·min1 (P = 0.2) respectively, although in the TT group significantly lower flows [377 (124) mL·min1], (P < 0.01) were achieved. Systolic pressure in the SLA groups was more stable in response to insertion than in the TT. With ProSealTM there was a lower incidence of sore throats than with TT (30% vs 57%), (P < 0.05), but there was a lesser difference as compared with SLIPATM (30% vs 49%), (P > 0.05). With both SLAs, there was a significant reduction in operating room time (> three minutes), (P < 0.001).
Conclusions: These results suggest that the ProSeal LMATM (reusable) and SLIPATM (single-use) SLAs were easy to use without requiring muscle relaxants, and reduce operating room time compared to the TT technique in day case laparoscopies.
| Introduction |
|---|
|
|
|---|
The Streamlined LIner of the Pharynx Airway (SLIPATM; SLIPAmed SA Pty Ltd, Cape Town, South Africa) is a new type of SLA, fabricated from soft plastic with an anatomically preformed shape that lines the pharynx. This means positive pressure ventilation may be achieved without a cuff inflating mechanism that is required in other airway types such as the laryngeal mask (LM). A full description of this device and its limitations is accessible on the SLIPA website (www.slipa.com). Briefly, it comprises a hollow, blow-moulded chamber shaped like a boot with toe (T), bridge (B) that seals at the base of the tongue and a heel (H), which anchors the device in a stable position between the esophagus and nasopharynx. It is not necessary for the SLIPATM to be tied or strapped into position (Figure 1
). The chamber provides a large capacity (50 mL compared to 3.5 mL in the LM) for providing maximum but limited storage of regurgitated liquids should they arise from the stomach, thus preventing their inadvertent overflow into the trachea.1 The safety advantages of simplicity and minimization of aspiration risk of a device without a cuff inflating mechanism necessitates double the number of sizes in order to obtain a good quality seal for positive pressure ventilation. The aspiration protection characteristics of both the SLIPATM and ProSealTM LM have been compared favourably with those of the LM in a laboratory study.1
|
| Materials and methods |
|---|
|
|
|---|
|
For all patients in whom successful airway placement was achieved, the lungs were ventilated mechanically through a circle system using a single Frontline 560 anesthetic machine (Blease Medical Equipment Ltd, Buckinghamshire, UK). The ventilator on this machine is time-cycled and volume-controlled with a rising bellows. The inspiration: expiration ratio was varied between 1:1 to 1:2 with a respiratory rate of 18 min1 in order to use the lowest inflation pressure to achieve adequate ventilation. The tidal volume was adjusted to achieve an end tidal CO2 value of 30 to 40 mmHg. Mild hyperventilation combined with the use of opioids meant that patients were not breathing spontaneously throughout the abdominal insufflation phase of the procedure. Systolic pressure was measured non-invasively before anesthesia, immediately following induction but before airway placement, and within two minutes of placing the airway. After surgery, women in the TT group received neostigmine and glycopyrrolate to reverse neuromuscular block. The time from start of deflation of the abdomen to the time the patient was ready to move to the recovery ward was recorded (recovery time). After surgery, all airways were removed in the operating room before transfer of the patient to the recovery room. The recovery room staff recorded the incidence of early sore throats. The number of late sore throats was recorded after a telephone survey on the first postoperative day. Even mild sensations in the throat with minimal pain were recorded as sore throats.
| Airway insertion times |
|---|
|
|
|---|
Quality of airway seal
In addition to recording peak airway inflation pressures, the quality of the airway seal was measured in two ways:
Recovery time
In order to exclude variations in the surgical times, the recovery time was measured from the beginning of abdominal deflation until the patient was deemed by the anesthesiologist to be ready for transfer to the recovery room with the airway removed.
Statistics
Distribution of baseline variables was assessed by the Kolmogorov-Smirnov and Shapiro-Wilk W tests. For each study, parameter summary statistics were tabulated. Differences in baseline variables between study groups were compared using two-tailed t tests or the Mann-Whitney U test for continuous data, and Chi-square with Yates continuity correction for discrete data. Comparison among study group was performed by analysis of variance (ANOVA). If ANOVA was significant, the differences among the three groups were assessed by the Student-Newman-Keuls test to allow for multiple comparisons. For non-normally-distributed continuous data, the Kruskal-Wallis test (non-parametric ANOVA) and Dunns test for multiple comparisons were employed. A P value < 0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
|
|
| Discussion |
|---|
|
|
|---|
We found that the TT sealed significantly better than both SLAs as demonstrated by the lower minimum flow rate into the absorber system. However, airway sealing pressures demonstrated SLA sealing qualities almost as good as those of a TT. The minimum flow rate required was well below 1 L·min1 thus allowing low flow anesthesia. The difference between the two SLAs was not significant. After abdominal insufflation of the abdomen with CO2, the lower inflation pressures noted when using the TT may well be due to the use of the muscle relaxant. When comparing the TT with the SLAs, the differences in the rise in systolic pressures in response to airway placement was of a clinically important magnitude (> 15%). However, the SLIPATM and PLMA are associated with similar hemodynamic responses.
Interestingly, satisfactory conditions for ventilation were consistently achieved without muscle relaxants using both SLAs. Recovery time was significantly lower when using either SLAs compared with the TT. The time saved using the SLAs most probably relates to the avoidance of the need for muscle relaxants and the associated time for reversal and extubation.1 The mean time saved in the operating room based only on recovery time was greater than three minutes. Given the average time taken for each procedure (Table I
), cumulative time saving could mean that, on average, one extra case per day could be performed in the day-surgery unit.
The combined incidence of both early and late sore throats showed that the PLMA caused the least sore throats. All sore throats recorded in the SLAs were mild or minimal in nature and required no treatment. There was only one moderate sore throat experienced by a patient in the TT group. The 9% loss to follow up related only to the 24-hr follow-up over the phone sore throat report. Considering that the patients were discharged the same day of the procedure (day surgery), we expected that we would be unable to contact some patients for the 24-hr follow-up. This is unlikely to be of clinical importance and, therefore, unlikely to have made an important difference to the final report.
The design of the PLMA is based upon the classical LMATM but with better maximum seal pressure, and it has a drainage tube, making it more suitable for laparoscopic work.6 However, it is not a single-use product. The SLIPATM is designed for single-use, and has a means of minimizing aspiration risk if regurgitation occurs.1 This relates to the large capacity within its hollow structure for entrapment of regurgitated liquids. Storage volumes (50mL)1 before aspiration occurred in the lung model were almost double the volume of the contents in fasted patients stomachs (26 mL).7,8 The effectiveness of the mechanism during a laparoscopy in the clinical part of the study was demonstrated in one patient who regurgitated 15 mL into the SLIPATM without any aspiration.1 As the reported incidence of aspiration with SLAs is very low, we can only know for certain that this proposed mechanism for preventing aspiration is reliable after it is widely used and perhaps after multicentre studies.
In conclusion, PLMA and SLIPATM SLA have been shown to be equally efficacious for use in gynecological laparoscopies without requiring muscle relaxants, with time-saving advantages and fewer side effects. There is less systolic pressure response to placement and fewer sore throats are experienced compared to airway management with a TT technique.
| Acknowledgments |
|---|
| Footnotes |
|---|
Accepted for publication August 19, 2005. Revision accepted September 15, 2005.
| References |
|---|
|
|
|---|
2 Williams MT, Rice I, Ewen SP, Elliott SM. A comparison of the effect of two anaesthetic techniques on surgical conditions during gynaecological laparoscopy. Anaesthesia 2003; 58: 5748.[Medline]
3 Piper SN, Triem JG, Rohm KD, Maleck WH, Schollhorn TA, Boldt J. ProSeal-laryngeal mask versus endotracheal intubation in patients undergoing gynaecologic laparoscopy. Anasthesiol Intensivmed Notfallmed Schmerzther 2004; 39: 1327.[Medline]
4 Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth 1999; 82: 2867.
5 Hein C, Plummer J, Owen H. Evaluation of the SLIPATM (Streamlined LIner of the Pharynx Airway), a single use supraglottic airway device, in 60 anaesthetised patients undergoing minor surgical procedures. Anaesth Intensive Care 2005; 33: 5717.[Medline]
6 Maltby JR, Beriault MT, Watson NC, Liepert D, Fick GH. The LMA-ProSeal is an effective alternative to tracheal intubation for laparoscopic cholecystectomy. Can J Anesth 2002; 49: 85762.
7 Juvin P, Fevre G, Merouche M, Vallot T, Desmonts JM. Gastric residue is not more copious in obese patients. Anesth Analg 2001; 93: 16212.
8 Maltby JR, Pytka S, Watson NC, McTaggart Cowan RA, Fick GH. Drinking 300 mL of clear fluid two hours before surgery has no effect on gastric fluid volume and pH in fasting and non-fasting obese patients. Can J Anesth 2004; 51: 1115.
This article has been cited by other articles:
![]() |
H. Hooshangi and D. T. Wong Brief Review: The Cobra Perilaryngeal Airway (CobraPLA(R)) and the Streamlined Liner of Pharyngeal Airway (SLIPATM) supraglottic airways: [Article de synthese court : Les dispositifs supraglottiques Cobra Perilaryngeal Airway (CobraPLA(R)) et Streamlined Liner of Pharyngeal Airway (SLIPATM)] Can J Anesth, March 1, 2008; 55(3): 177 - 185. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Miller Laryngeal Mask Airway Classic and Streamlined Pharynx Airway Liner Airway Comparison Anesth. Analg., November 1, 2007; 105(5): 1508 - 1508. [Full Text] [PDF] |
||||
![]() |
M. Lange, T. Smul, P. Zimmermann, R. Kohlenberger, N. Roewer, and F. Kehl The Effectiveness and Patient Comfort of the Novel Streamlined Pharynx Airway Liner (SLIPA(R)) Compared with the Conventional Laryngeal Mask Airway in Ophthalmic Surgery Anesth. Analg., February 1, 2007; 104(2): 431 - 434. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |