| Abstract |
|---|
|
|
|---|
Methods: Forty patients scheduled for general anesthesia requiring endotracheal intubation were studied prospectively. Preoperatively, the patients airways were evaluated according to Mallampati score, thyromental distance and interincisor gap. After induction of anesthesia laryngoscopy was performed with the original laryngoscope and its modified counterpart in random order. A lateral x-ray of the neck was taken after the optimal view had been obtained, and blade-tooth distance, laryngeal view, blade-tooth contact and need for assistance were measured. Using angular calculations the laryngoscopes were analyzed at different insertion depths on graph paper, and the results were compared with data from the lateral x-rays.
Results: With a modified Macintosh blade the blade-tooth distance was significantly greater in comparison to the original design (2.5 ± 2.1 cm vs 0.2 ± 0.1 cm, P < 0.01). Consequently the number of blade-tooth contacts was significantly lower (20% vs 75%, P < 0.05). The best laryngeal view could be obtained using a modified Macintosh laryngoscope. With a modified Miller laryngoscope laryngeal visibility was not improved and assistance was required more often to achieve adequate intubating conditions (35% vs 5%, P < 0.05).
Conclusion: A reduction of the proximal flange of a Miller blade decreases the blades effectiveness for laryngeal visualization, whereas a similar modification of a Macintosh blade increases blade-tooth distance, decreases the number of blade-tooth contacts and provides a better laryngeal view.
| Introduction |
|---|
|
|
|---|
|
| Methods |
|---|
|
|
|---|
A computer-generated random number table was used to randomize patients in two steps: first for the type of laryngoscope, and second for blade sequence. A sealed envelope method was used for allocation concealment. Patients were randomized accordingly into one of four groups: 1) Macintosh #4, modified Macintosh #4; 2) modified Macintosh #4, Macintosh #4; 3) Miller #4, modified Miller #4; 4) modified Miller #4, Miller #4.
In the operating room, the patients head was placed on a pillow to achieve the anatomic sniffing position, after which routine monitors were applied and iv access was established. Anesthesia was induced with propofol 23 mg·kg1 iv, remifentanil infused at a rate of 1 µg·kg1·min1 over two minutes, and cisatracurium 0.15 mg·kg1 iv. Following induction, all laryngoscopies were performed by one experienced anesthesiologist, accustomed to both Macintosh and Miller blades. During the study period anesthesia was maintained with infusions of propofol (8 mg·kg1·hr1) and remifentanil (0.5 µg·kg1·min1). Adequacy of muscle relaxation was assessed by acceleromyography (TOF-Watch, Organon; West Orange, NJ, USA), and depth of anesthesia was measured with a bispectral index (BIS) XP monitor (Aspect Medical Systems; Newton, MA, USA). In all patients BIS values remained < 55 throughout the study period. The laryngoscope was introduced at the right side of the mouth and moved towards the midline, shifting the tongue leftwards. With the epiglottis held anteriorly (Miller blade) or posteriorly (Macintosh blade), optimum visibility of the glottis was established. Once the best laryngeal view had been obtained, a research assistant measured the perpendicular tooth-to-heel distance from the tip of the upper incisors to the heel of the laryngoscope. The visibility as obtained by the laryngoscopist was graded according to Cormack and Lehane (1: glottis visible; 2: only posterior part of glottis visible; 3: only epiglottis visible; 4: no laryngeal structures visible). A lateral x-ray of the neck was taken, and the laryngoscope was removed thereafter. If SpO2 decreased to a value < 90% the patient was ventilated immediately via face mask with 100% oxygen. Only after restoration of oxygen saturation would a second laryngoscopy be attempted. Assistance during laryngoscopy was limited to counteracting anteflexion of the head when anteflexion impaired laryngeal view, or retracting the upper lip, when it obstructed the laryngeal view. Any blade-tooth contact was recorded. As soon as all measurements had been recorded and x-rays were completed, the patients trachea was intubated, and the study protocol was considered complete. For intubation there were no restrictions with regard to type of laryngoscope or extent of assistance. The remainder of intraoperative and postoperative care was at the discretion of the attending anesthesiologist.
The angular parameters of the laryngoscope blades for different insertion depths were calculated according to the method of Marks et al.10 with photographs of the laryngoscope blades placed on Cartesian graph paper. In patients the angular parameters were derived from x-ray laryngoscopies taken when optimal laryngeal visibility had been achieved. These parameters were calculated as follows (Figure 2
): J is a point between the mandibular condyles; S is in the middle of the symphysis menti; I is the point where the incisors touch the laryngoscope; T is the tip of the laryngoscope, where it ideally touches the hyoid bone. The line IT corresponds to the ideal line of view and represents the reference line. The point M lies on the blade surface at the level of the mandible on the line JS. If the point M lies in front of the line IT, the angle MIT has a positive value and signifies forward space encroachment. If M lies behind the line IT, MIT has a negative value and signifies space enhancement. Thus, MIT provides a measure of how closely the laryngoscope can move towards the mandible, a more negative MIT angle indicates better laryngoscope performance. A measure for the eyeline deviation is derived from the angle EIT, with the line IE being the actual line of view and IT being the optimal line of view. A smaller angle (typically for straight blades) formed by the lines IE and IT implies less deviation from the ideal line of view. A combination of both scores (EIT + MIT) offers the possibility to compare different laryngoscopes at different insertion depths on a theoretical basis. A lower combination score signifies a better laryngoscope performance. The angles derived from the radiographs, determined from the point of optimal laryngeal visibility, were calculated with a high resolution system (accuracy ± 0.1°) by an experienced radiologist.
|
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Reducing the proximal part of the flange of a Macintosh laryngoscope resulted in a lower frequency of laryngoscope-tooth contacts. In accordance with other studies9,1114 the distance between flange and teeth was increased significantly, while the visibility of laryngeal structures was improved in comparison to the original Macintosh blade. According to Lockhart et al.,15 51% of dental traumas in anesthesiology occur during tracheal intubation by contact of the laryngoscopes flange with the maxillary left incisor. An increased blade-tooth distance and less blade-tooth contact can therefore be assumed to be associated with a decreased risk of dental injury. In accordance with Marks et al.10 and Yardeni et al.16 the theoretically derived angular parameters of the modified Macintosh blade are reliable estimations for the actual measurements in patients. The better combination score of forward space encroachment and eyeline deviation correctly predicts a better laryngoscope performance for the modified Macintosh blade.
The modification of the Miller blade did not improve the blades performance. No improvement of laryngeal view or increase of laryngoscope-tooth distance was observed. Our experienced anesthesiologist had problems intubating the trachea with the modified Miller blade in seven cases without assistance. Assistance during laryngoscopy consisted of counteracting anteflexion of the head when anteflexion impaired laryngeal view, or retraction of the upper lip, when it obstructed the laryngeal view.
A pronounced tendency towards anteflexion was observed with both modified laryngoscope blades. This finding was also observed by Bucx et al.9 and is most likely the result of different relations of torque and force for these modified blades.
This study had several limitations. First, it was not possible to evaluate all four laryngoscope blades in each patient, due to unacceptable radiation exposure and laryngeal tissue stress. Moreover, in this study we focused upon optimal laryngoscopic views, yet it has to be emphasized that a good laryngoscopic view does not always equate with easy tracheal intubation. The patient was intubated after the second x-ray. Successful intubation of the trachea was not one of the study endpoints, as it would not have been justifiable to intubate the patients trachea twice. Another limitation is that a longer teeth flange distance has not yet been proven to decrease the incidence of dental damage. Other methods to decrease dental injury in at-risk patients include tooth protectors,17 angulated blades,18 blades with soft heels or with no heel at all.6 So far, no single approach has proven to be superior with respect to the prevention of tooth damage. Although larger controlled trials, comparing different laryngoscopes with regard to laryngeal visibility19 or frequency of blade-tooth contact11 have been conducted, trials with dental injury damage as a primary outcome parameter would be ethically challenging, and are still lacking. Finally, the majority of patients in this study did not feature any airway abnormalities, and our results may not be applicable to patients with difficult airways. Further studies with laryngoscope blades with a Callander modification focusing on this patient group, may be warranted.
In conclusion, a Callander modification of the Miller laryngoscope blade is associated with significant disadvantages, requiring increased need for assistance in airway management and no improvement of laryngeal vision. In contrast, a Callander modification of the Macintosh laryngoscope blade with a reduced flange at the proximal end provides more distance between maxillary incisors and the laryngoscope, and an improved laryngeal view. This modification of the Macintosh blade decreases the frequency of contact between blade and incisors, and may potentially decrease the likelihood of dental trauma.
| Footnotes |
|---|
Presented in part at the Annual Meeting of the European Society of Anesthesiologists, Lisbon, June 2004.
Competing interests: None declared.
Accepted for publication December 18, 2005. Revision accepted January 20, 2006. Final revision accepted January 30, 2005.
| References |
|---|
|
|
|---|
2 Dorsch JE, Dorsch SE. Laryngoscopes. In: Dorsch JE, Dosch SE (Eds). Understanding Anesthesia Equipment, 3rd ed. Baltimore: Williams and Wilkins; 1984: 399437.
3 Cooper SD. Laryngoscope blades and lighted stylets. In: Hanowell LH, Waldron RJ, Hwang JCF (Eds). Airway Management, 1st ed. Philadelphia: Lippincott-Raven; 1996: 22357.
4 Owen H, Waddell-Smith I. Dental trauma associated with anaesthesia. Anaesth Intensive Care 2000; 28: 13345.[Medline]
5 Warner ME, Benenfeld SM, Warner MA, Schroeder DR, Maxson PM. Perianesthetic dental injuries: frequency, outcomes and risk factors. Anesthesiology 1999; 90: 13025.[Medline]
6 Bizzarri DV, Giuffrida JG. Improved laryngoscope blade designed for ease of manipulation and reduction of trauma. Anesth Analg 1985; 37: 2312
7 Ibler M. Modification of Macintosh laryngoscope blade. Anesthesiology 1983; 58: 2001.[Medline]
8 Callander CC, Thomas J. Modification of Macintosh laryngoscope for difficult intubtion. Anesthesia 1987; 42: 6712.
9 Bucx MJ, Snijders CJ, ven der Vegt MH, Holstein JD, Stijnen T. Reshaping the Macintosh blade using biomechanical modelling. A prospective comparative study in patients. Anaesthesia 1997; 52: 6627.[Medline]
10 Marks RR, Hancock R, Charters P. An analysis of laryngoscope blade shape and design: new criteria for laryngoscope evaluation. Can J Anaesth 1993; 40: 26270.[Abstract]
11 Lee J, Choi JH, LeeYK, Kim ES, Kwon OK, Hastings RH. The Callander laryngoscope blade modification is associated with a decreased risk of dental contact. Can J Anesth 2004; 51: 1814.
12 Watanabe S, Suga A, Asakura N, et al. Determination of the distance between the laryngoscope blade and the upper incisors during direct laryngoscopy: comparisons of a curved, an angulated straight, and two straight blades. Anesth Analg 1994; 79: 63841.
13 Maleck WH, Koetter KP, Lenz M, Piper SN, Triem J, Boldt J. A randomised comparison of three laryngoscopes with the Macintosh. Resuscitation 1999; 42: 2415.[Medline]
14 Papageorgiou E, Kokkinis K, Goumas P, Mochloulis G, Alexopoulos C. Objective evaluation of the difficulty of endotracheal intubation. A comparative study of two different laryngoscope blades. Anaesth Intensive Care 1997; 25: 6558.[Medline]
15 Lockhart PB, Feldbau EV, Gabel RA, Connolly SF, Silversin JB. Dental complications during and after tracheal intubation. J Am Dent Assoc 1986; 112: 4803.[Abstract]
16 Yardeni IZ, Gefen A, Smolyarenko V, Zeidel A, Beilin B. Design evaluation of commonly used rigid and levering laryngoscope blades. Acta Anaesthesiol Scand 2002; 46: 10039.[Medline]
17 Nakahashi K, Yamamoto K, Tsuzuki M, et al. Effect of teeth protector on dental injuries during general anesthesia (Japanese). Masui 2003; 52: 2631.[Medline]
18 Bellhouse CP. An angulated laryngoscope for routine and difficult tracheal intubation. Anesthesiology 1988; 69: 1269.[Medline]
19 Arino JJ, Velasco JM, Gasco C, Lopez-Timoneda F. Straight blades improve visualization of the larynx while curved blades increase ease of intubation: a comparison of the Macintosh, Miller, McCoy, Belscope and Lee-Fiberview blades. Can J Anesth 2003; 50: 5016.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||