| Abstract |
|---|
|
|
|---|
Clinical features: An 81-yr-old patient suffered from an intravascular knotting of a Seldinger wire during subclavian venous cannulation. We describe a technique for successful removal of knotted guidewire under fluoroscopic guidance using the vessel dilator of a central venous cannulation kit. In this case, the technique was successful without associated immediate or delayed complications. Although central venous cannulation with the Seldinger technique is a commonly performed procedure, it may result in numerous complications, including kinking, and rarely complete knotting of the guidewire.
Conclusions: A thorough understanding of procedural complications and physical characteristics of the guidewire is vital in order to ensure patient safety when using the Seldinger technique for central venous cannulation. We have reviewed the relevant literature for guidewire design and structure, associated complications, and provide recommendations for safe use of guidewires.
| Introduction |
|---|
|
|
|---|
| Case report |
|---|
|
|
|---|
On the day of admission his blood pressure was 145/81 mmHg with a pulse rate of 76 beats·min1. He was 175 cm tall and had a body mass index of 32 kg·m2. He underwent total gastrectomy under general anesthesia, which was uneventful. A feeding jejunostomy tube was sited intraoperatively. The patient made a good immediate postoperative recovery without significant problems.
Later he developed severe postoperative ileus. On the ninth postoperative day a computed tomography scan showed dilated intestine with no other significant findings. It was decided that he should have central venous access to facilitate total parenteral nutrition. He was referred to the anesthetic team for insertion of a central venous catheter prior to discharge from the high dependency unit (HDU).
Verbal consent for central venous cannulation via the subclavian approach was obtained. Standard barrier protection was maintained, and 7 mL of 1% lidocaine was injected to the entry site in the right infraclavicular area. In line with standard practice, a Seldinger wire technique was used to site a 7 Fr triple lumen catheter. The subclavian vein was identified on the first attempt without complications. However, threading the 0.81 mm J-tip guidewire through the 18 G needle proved difficult. Needle and wire were removed from the insertion site en bloc. At the second attempt using the same insertion site, the vein was re-identified uneventfully. This time the Seldinger wire was passed through the needle, though a slight resistance was noted. After removal of the needle, the dilator was threaded over the wire. On removal of the dilator the wire was found to be kinked, so it was decided to remove the wire. The wire was initially withdrawn easily, but at approximately 7 cm remaining in situ further removal was not possible despite firm tension. Consultant help was sought and the tract was dissected under local anesthesia, however any further attempts at removal were unsuccessful. General surgical advice was sought while a chest x-ray was performed, which showed the guidewire to be looped and knotted in the periclavicular area (Figure 1
).
|
|
| Discussion |
|---|
|
|
|---|
The length and diameter of these wires are variable to meet specific needs. In older versions many spring guidewires also incorporated a smaller straight wire (safety wire) running parallel to the inner core (mandrel), which was attached to both ends of the coil. This smaller core wire was employed to reduce the possibility of kinking or stretching the coil with resultant fracture. In addition, it prevented fragment migration or embolization if coil separation occurred.7
The specific design of these wires is a patent of each manufacturer. Edwards Life Sciences have informed us that their guidewires comply with ISO 11070:1998 standards (personal communication). However the part of the standardA relevant to coated guidewires states:
"The guide wire to be coiled several times around a cylinder whose radius is scaled to the diameter of the guide wire".
The guidewire then is examined visually for the presence of cracks, splits, and dehiscence.8
However, compliance with these standards in North America does not imply that the guidewire meets FDA standards.9 The stability of the guidewire is routinely examined at certain tension loads. Monaca et al.6 have described some other details of guidewire manufacturing and testing. According to these, the weld seams between the inner core and the outer covering at both ends are supposed to withstand tension of at least 17 N (1.73 kg force) before they detach, whereas in its middle section, the complete guidewire should withstand a stress of approximately 290 N (29.57 kg force).
The outer material, the coil, of the guidewire is usually made of stainless steel and the inner material is made of nitinol, which is an alloy of nickel and titanium (personal communication, B. Braun). Recently, a new type of guidewire has been developed at Tohoku University in Japan, which is a Cu-Al-Mn (copper, aluminium and manganese) alloy. This has graded characteristics with super elasticity at the tip as well as high stiffness in the middle part. This allows greater tension and rotational stress to be applied.10 This new guidewire is not yet available in commercial central venous catheter kits.
The Seldinger technique is not devoid of complications attributable purely to the employment of guidewires. Extravascular3,11 and intravascular46 looping and knotting are well-reported complications. Shearing12 and fracture, or impaction of the wires have also been reported. Similarly, paradoxical embolization of a fractured guidewire13 is a known complication. These complications are related to damage or mechanical disruption of the normal structure of the wire.
However, occasionally even the insertion of an entirely intact guidewire can cause problems. Excess introduction of the wire may result in cardiac arrhythmias. 14 In one case, excessive introduction of a guidewire caused a cyst arising from the tricuspid valve annulus to prolapse into the right ventricular outflow tract and cause cardiac arrest with electromechanical dissociation.15 Contamination16 of the guidewire during insertion has been reported in the literature as well as misplacement and loss of guidewires in the cardiovascular system.17 Special precautions are required in patients with inferior vena cava filters since there have been numerous reports of entrapment of guidewire in these filters.18,19 There have been rare cases of fatal pericardial tamponade20,21 following guidewire insertion.
Intravascular and extravascular looping and knotting and any further damage to the structure of the guidewires can be prevented if caution is used during their handling. Guidewires are not very rigid structures and if any force is applied they are likely to kink or loop. Further application of force after looping might result in knot formation. There are two elements in the sequence of the Seldinger technique that are potentially important in preventing the kinking of the wire. First, during insertion of the wire through the needle no resistance should be encountered, and the wire should pass freely, without any force into the vein. To avoid shearing, the wire should not be pulled back through the needle once it has passed the bevel. Second, care should be exercised when passing the dilator over the wire. Forcing the dilator in a direction that diverges from the path of the wire can result in kinking the wire and potentially cutting out through the vein with possibly fatal consequences.22
Knotting of the wires is almost exclusively described in the literature following the subclavian approach, although it has also been described once during attempted femoral vein cannulation.23 This could be due to the fact that the path of the subclavian vein is not straight, as it loops over the first rib to descend into the superior mediastinum. Further to this, there is a pinch area between the clavicle and the first rib, which can impede the anterograde threading of the wire and might contribute to looping and knotting.
The axillary vein approach is an alternative to the subclavian approach. Attractions of this technique include the ease with which the axillary vein can be visualized using ultrasound.24 This technique is described as having a low complication rate although it is uncertain whether the incidence of guidewire complications per se is lower when cannulation is ultrasound-guided.25 We suggest that the following precautions may reduce the risk of complications. First, special care should be taken in known cases of altered anatomy or where previous cannulation has occurred at the same site. On encountering any resistance the guidewires should not be advanced. The guidewire and needle should be withdrawn as an assembly to prevent the cutting and shearing of the wire by the sharp needle tip. After more than one attempt, the integrity of the wire should be rechecked. A sufficiently large skin incision should be used in order to thread the dilator and the line over the wire smoothly to prevent kinking of the wire at the skin. One end of the guidewire should always be held by the operator at all times to minimize the risk of loss in the blood vessel. Some manufacturers allow use of the straight end of the wire on insertion; but in general this is not recommended.B If the wire is fixed, force should not be applied to remove it. This might result in separation of the coil and the core resulting in fracture of the wire. Finally, if minimal to moderate force needs to be applied then forceps should be used to grasp the wire in a way that the outer coil and the inner core are grasped together and separation does not occur.
In conclusion, although it is not possible to study the optimal method of removal of a knotted guidewire, there are case reports suggesting that this should be done under fluoroscopic guidance. A vessel dilator can be used to tighten the knot into its smallest circumference, allowing it to be withdrawn along with the dilator. Closure of the vessel in a controlled fashion, for example by a purse-string suture,26 may limit the risk of hemorrhage and air embolism, and is best performed by a vascular surgeon in the operating room, as in our case. We suggest that this approach should be followed, as removal of a looped or knotted wire may cause vessel damage, major hemorrhage and the need for urgent surgical intervention. Adhering to the recommended precautions will reduce the risk of mechanical complications associated with the Seldinger technique.
| Footnotes |
|---|
Accepted for publication January 12, 2007. Revision accepted February 19, 2007.
A ISO. Sterile, single-use intravascular catheter introducers. Geneva, Switzerland. International Organization for Standardization. ISO 11070:1998. ![]()
B B. Braun. Manufacturers instructions for use. Central venous catheter for Infusion therapy. Melsungen AG, Melsungen, Germany: February 2002. ![]()
| References |
|---|
|
|
|---|
2 Mansfield PF, Hohn DC, Fornage BD, Gregurich MA, Ota DM. Complications and failures of subclavian-vein catheterization. N Engl J Med 1994; 331: 17358.
3 Wang LP, Einarsson E. A complication of subclavian vein catheterisation. Extravascular knotting of a guidewire. Acta Anaesthesiol Scand 1987; 31: 1878.[Medline]
4 Carpentier JP, Braz da Silva J, Choukroun G. Formation of a knot in a J spiral metallic guide: a complication of the Seldinger method (French). Cah Anesthesiol 1991; 39: 2778.[Medline]
5 Koundouris C, Tornaris G, Charopoulos I, Vounasis A, Solomou G. Subclavian vein flexible guidewire knotting. A potential serious complication in hemodialysis patients. Chirurgia (Bucur) 2004; 99: 614.[Medline]
6 Monaca E, Trojan S, Lynch J, Doehn M, Wappler F. Broken guide wire - a fault of design? Can J Anesth 2005; 52: 8014.
7 Schwartz AJ, Horrow JC, Jobes DR, Ellison N. Guide wires--a caution. Crit Care Med 1981; 9: 3478.[Medline]
8 Triolo P. Using risk analysis to develop coated medical devices. Available from URL; http://www.devicelink.com/mddi/archive/05/01/006.html (accessed 28/05/06) 2005.
9 U.S. Food and Drug Administration. Recognition and use of consensus standards; final guidance for industry and FDA staff, (Rockville, MD: FDA:); Available from URL; http://www.fda.gov/cdrh/ost/guidance/321.html (accessed 28/05/06) CDRH, June 2001.
10 Sutou Y, Omori T, Furukawa A, et al. Development of medical guide wire of Cu-Al-Mn-base superelastic alloy with functionally graded characteristics. J Biomed Mater Res B Appl Biomater 2004; 69: 649.[Medline]
11 Hoogsteden L, Filshie J, Querci della Rovere G. Knotted guidewire A complication of Hickman line insertion. Anaesthesia 1996; 51: 713.[Medline]
12 Chhabra B, Kiran S, Sekhri C. Shearing of guide wire: a complication of central venous catheterization. Journal of Anaesthesiology 2002; 18: 99101.
13 Casserly IP, Goldstein JA, Rogers JH, Lasala JM. Paradoxical embolization of a fractured guidewire: successful retrieval from left atrium using a snare device. Catheter Cardiovasc Interv 2002; 57: 348.[Medline]
14 Lee TY, Sung CS, Chu YC, Liou JT, Lui PW. Incidence and risk factors of guidewire-induced arrhythmia during internal jugular venous catheterization: comparison of marked and plain J-wires. J Clin Anesth 1996; 8: 34851.[Medline]
15 Quiney NF. Sudden death after central venous cannulation. Can J Anaesth 1994; 41: 5135.
16 Palmer S, Solano T. Contamination of guidewires during insertion of central venous catheters in an intensive care setting. Infect Control Hosp Epidemiol 2005; 26: 5067.[Medline]
17 Lum TE, Fairbanks RJ, Pennington EC, Zwemer FL. Profiles in patient safety: misplaced femoral line guidewire and multiple failures to detect the foreign body on chest radiography. Acad Emerg Med 2005; 12: 65862.[Medline]
18 Duong MH, Jensen WA, Kirsch CM, Wehner JH, Kagawa FT. An unusual complication during central venous catheter placement. J Clin Anesth 2001; 13: 1312.[Medline]
19 Liddell RP, Spinosa DJ, Matsumoto AH, Angle JF, Hagspiel KD. Guidewire entrapment in a Greenfield IVC filter: rail and reins technique. Clin Radiol 2000; 55: 87881.[Medline]
20 Cavatorta F, Campisi S, Fiorini F. Fatal pericardial tamponade by a guide wire during jugular catheter insertion. Nephron 1998; 79: 352.[Medline]
21 Booth SA, Norton B, Mulvey DA. Central venous catheterization and fatal cardiac tamponade. Br J Anaesth 2001; 87: 298302.
22 Jankovic Z, Boon A, Prasad R. Fatal haemothorax following large-bore percutaneous cannulation before liver transplantation. Br J Anaesth 2005; 95: 4726.
23 Wong CS, Wong KL, Sit KF, Chen CC, Sia SL, Wei TT. Guide wire knotting in the femoral vein during central venous catheterization (Chinese). Ma Zui Xue Za Zhi 1992; 30: 557.[Medline]
24 Galloway S, Bodenham A. Ultrasound imaging of the axillary vein--anatomical basis for central venous access. Br J Anaesth 2003; 90: 58995.
25 Sharma A, Bodenham AR, Mallick A. Ultrasound-guided infraclavicular axillary vein cannulation for central venous access. Br J Anaesth 2004; 93: 18892.
26 Pachai A. Guidewire knot formation during insertion of a central venous catheter (Danish). Ugeskr Laeger 1997; 159: 55345.[Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |